Auto Use this mode to let the acquisition free-run in the
absence of a valid trigger; This mode allows an
untriggered, scanning waveform at 100 ms/div or
slower time base settings.
Normal Use this mode when you want to see only valid
triggered waveforms; when you use this mode, the
oscilloscope does not display a waveform until after
the first trigger.
Mode
Single when you want the oscilloscope to acquire a single
waveform, press the “SINGLE ”button.
Set up Enter the “Trigger Setup Menu”(See table 2-14).
Table 2-14 Trigger Setup function menu
Option Setting Explain
DC Passes all components of the signal
AC
Blocks DC components and attenuates signals
below 10 Hz.
HF Reject
attenuates the high-frequency components above 80
kHz.
Coupling
LF Reject
Blocks the DC component and attenuates the
low-frequency components below 300 kHz.
Holdoff Using the “universal” knob to adjust holdoff
time(sec)ˈthe holdoff value is displayed.
Holdoff
Reset
Reset holdoff time to 100ns.
Return Return the first page of “ Trigger main menu”.
Picture 2-30
Operate Instruction:
1. Set up Type
1) Press the “TRIG MENU” button to display “Trigger” menu.
24