User Manual

Note˖The oscilloscope can capture a waveform in normal YT mode
at any sampling rate. You can view the same waveform in XY mode.
To do so, stop the acquisition and change the display format to XY.
Operation steps ˖
Ɣ Channel 1“Volt/div”and vertical “POSITION” set up the horizontal scale
and position.
Ɣ Channel 2“Volt/div”and vertical “POSITION” set up the horizontal scale
and position.
Ɣ Turn the “S/div” knob to adjust the sampling rate
Ɣ The following functions are forbidden at XY display form:
Ɣ Benchmark wave form and wave mathematic
Ɣ Cursor
Ɣ Auto˄resets display format to YT˅
Ɣ Trigger Control
Ɣ Horizontal Position Knob
Ɣ Vector Display Type
Ɣ Scan Display
Note˖When the display Type is set to Vectors, the oscilloscope
connects the sample points by using digital interpolation.
Interpolation mode contain linearity interpolation and sin(x)/x
interpolation. Sin(x)/x interpolation is suitable for real time mode.
It is effective at 250ns/div or at faster time base in real time mode.
65