Instruction manual
Introduction and Specifications
1-8
1720/1721
Table 1- 8: Certifications and Compliances (cont.)
Category Standards or description
Australia/New Zeal and
Declaration of Conformity -- EMC
Complies with EMC provision of Radiocommunications Act per the following standard(s):
AS/NZS 2064.1/2 Industrial, Scienti fic, and Medical Equipment: 1992
AS/NZS 3548 Inform ation Technology Equipment: 1995
FCC Compliance Emissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits.
Installation (Overvoltage)
Category
Terminal s on this product may have different installation (overvoltage) category designations. The
installation categories are:
CAT II I Distribution-level mai ns (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.
CAT II Local-level mains (wall sockets). Equipment at this level incl udes appli ances, portabl e
tools, and similar products. Equipment is usually cord-connected.
CAT I Secondary (signal level) or battery operated circuits of electronic equipment.
Pollution Degree A measure of the contaminates that could occur in the environment around and within a product.
Typical ly the internal environment inside a product is considered to be the same as the external.
Products shoul d be used only in the environment for which they are rated.
Pollution Degree 2 Normally only dry, nonconductive pollution occurs. Occasi onally a
temporary conduct ivity that is caused by condensation must be
expected. This location is a typical office/home environment.
Temporary condensation occurs only when the product is out of
service.
EC Declarat ion of Conformi ty --
Low Vol tage
Compliance was demonstrated to the following specification as listed in the Official Journal of the
European Union:
Low Vol tage Directive 73/ 23/EEC, amended by 93/68/EEC
EN 61010-1 Safety requirements for electrical equipment for measurement
control and laboratory use.
U.S. Nati onally Recognized
Testi ng Laboratory Listing
UL1244 Standard f or electrical and electronic measuring and testing
equipment.
Canadian Cert ific ation CAN/CSA C22.2 No. 231 CSA safety requirements for electrical and electronic measuring and
test equipm ent.
Additional Compliance IEC61010-1 Safety requirements for electrical equipment for measurement,
control, and laboratory use.
Installation (Overvoltage)
Category
CAT II As defined in IEC 1010--1, Annex J.
Pollution Degree Pollution Degree 2 As def ined in IEC 1010--1. Rated for indoor use only.
Safety Certification Compliance Temperature, operating: 0 to 50° C.
Altitude (maximum operating) 2,000 meters.
Equipment Type Test and measuring.
Safety Class Class I (as defined in IEC 1010--1, Annex H) -- grounded product.