User's Manual

Appendix J: All-In-One Computer User Manual
TM-5010-24P/24F User’s manual
10
interruptions
and
voltage
variations
on power
supply
input lines
IEC
61000-4-11
0% UT
(100 % dip in
UT)
for 0.5 cycle
0 % UT
(100 % dip in
UT)
for 1 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
0 % UT
(100 % dip in
UT)
for 250
cycles
0 % UT
(100 % dip in
UT)
for 0.5 cycle
0 % UT
(100 % dip in
UT)
for 1 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
0 % UT
(100 % dip in
UT)
for 250 cycles
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
model
TM-5010-24P/24F
requires continued
operation during
power
mains interruptions, it
is recommended that
the model
TM-5010-24P/24F be
powered from an
uninterruptible power
supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The model TM-5010-24P/24F is intended for use in the electromagnetic environment specified below.
The customer or the user of the model TM-5010-24P/24F should assure that it is used in such an
environment.
Immunity
IEC
60601
test level
Compliance
level
Electromagnetic environment guidance
I-23