User's Manual

6. xvision-Spine System Preprocess Workflow
6.2. Prepare Surgical Tools
Warning: The XVISION-SPINE system is not compatible with bent instruments. Using bent or
damaged instruments may affect system accuracy. Verify that the instrument is not
bent or otherwise damaged.
Universal Tool Adaptors
Figure 18: Prepare Surgical Tool for Universal Tool Adaptor
2
1
3
6
4
5
1 Remove the appropriate instrument marker from the sterile packaging and set it on the
universal tool adaptor.
2 Use the hex screwdriver to tighten the instrument marker to the universal tool adaptor.
3 Insert the surgical instrument.
4 For swivel tools, press the locking button.
5 Manually tighten the universal tool adaptor to the instrument.
6 Then use the XVS wrench to complete tightening the universal tool adaptor to the
instrument.
6-5