User's Manual

6. xvision-Spine System Preprocess Workflow
6.2. Prepare Surgical Tools
VP Tool Adaptors
Figure 20: Prepare Surgical Tool for VP Tool Adaptors
1
2
3
4
1 Remove the appropriate instrument marker from the sterile packaging and set it on the vp
tool adaptor.
2 Use the hex screwdriver to tighten the instrument marker to the vp tool adaptor.
3 Align the surgical instrument.
4 Tighten the vp tool adaptor to the instrument using a 3 mm Allen key (not supplied).
For all tool adaptors:
Warning: Make sure that the adaptor is firmly attached to the surgical instrument and cannot
be moved up or down. A change in the height of the after completing calibration
may affect the accurate display of the instrument’s tip.
Caution: Do not overtighten.
6-7