User's Manual

6. xvision-Spine System Preprocess Workflow
6.5. Patient Registration Using Spine Clamp
1 Attach the Z-Marker to the patient clamp with the allen screwdriver. Choose either the Z-
Marker-60 or Z-Marker-40 so it will be as close as possible to the surface of the patient’s
body.
Note: Make sure that the Z-Marker does not touch the patient’s skin.
2 Initiate an intraoperative 3D scan of the area, verifying that the desired spine anatomy and
the Z-Marker are included in the scan. The Z marker needs to be seen in the A-P and
lateral view before initiating a scan.
3 Remove the Z-Marker and set it aside for cleaning and sterilization. See Cleaning Reusable
Components on page8-1 for more information.
4 Attach the patient marker to the patient clamp with the allen screwdriver.
Continue with Spine Surgery with the xvision-Spine System (page 7-1).
6-17