Use Instructions

Troubleshooting
Problem
Possible Solution
Cannot power on the Aulisa
Oximeter Module
1. Make sure the Aulisa Oximeter Module is
kept away from any magnetic devices while
using.
2. Fully charge the Aulisa Oximeter Module.
Unable to obtain a valid SpO
2
or pulse rate reading
NOTE: In some instances,
perfusion of person being
monitored may be inadequate
for pulse detection.
1. Reposition the Aulisa Oximeter Sensor(s)
and keep it motionless for at least 10
seconds.
2. Position the Aulisa Oximeter Sensor(s) to a
different site.
3. Make sure the Aulisa Oximeter Module is
attached to the wrist and finger securely.
4. Check the accessories for any visible signs
of deterioration.
5. Warm the finger by rubbing or covering
with a sock.
6. Allow the foot to rest comfortably without
squeezing or pressing the sensor probe on a
hard surface.
7. Make sure the Aulisa Oximeter Module is
within 32.8 feet (10 meters) spherical
radius to the Receiver/Transponder.
8. Reduce or eliminate any interference. Make
sure the Aulisa Oximeter Module is NOT
placed on the same arm being used for
other medical therapies or diagnostics (e.g.
blood pressure cuff).
9. Check the Aulisa Digital Vital Sign
Monitoring System for any alarms or error
messages.
10. Check if the Aulisa Oximeter Module is in
low power.
11. Verify the system connection with the
Aulisa Digital Vital Sign Monitoring System.
12. Make sure that the Aulisa Oximeter Module
is not in proximity with other RF radiating
devices (such as diathermy, electrocautery,
RFID, and security systems).
Unstable or constant SpO
2
and
1. Shield the sensor probe from any light