Use Instructions

Pulse Rate Accuracy
Pulse rate accuracy has been functionally tested against an electronic pulse simulator at 30, 50,
80, 100, 150, 200, 250, and 290 bpm, with combinations of pulse amplitude settings of 0.5, 1, 3,
5, 7, 10, 13, 15, 17 and 20, and SpO
2
settings of 100%, 90%, 80%, 70%, 60%, 50%, 40%, 30%,
20%, 10%, and 1%. All 880 combinations of testing points (=8 x 10 x 11) of pulse rate passed the
± 3% acceptance criteria.
Equipment Response Time
This device uses a moving average to determine the pulse rate and SpO
2
. The following table
shows the equipment response time of this device.
Equipment Delays (second)
Data Averaging
≤ 4 seconds
Alarm Condition Delay
≤ 4 seconds
Alarm Signal Generation Delay
0 second
Data Update Period
1 second