NUFLOTM MC-III™ EXP Flow Analyzer User Manual Manual No. 9A-50165003, Rev.
Important Safety Information Symbols and Terms Used in this Manual ! WARNING: This symbol identifies information about practices or circumstances that can lead to personal injury or death, property damage, or economic loss. CAUTION: Indicates actions or procedures which if not performed correctly may lead to personal injury or incorrect function of the instrument or connected equipment.
MC-III™ EXP Flow Analyzer Table of Contents Contents Important Safety Information................................................................................................................................ii Section 1—Introduction.................................................................................................................................... 7 Operation....................................................................................................................................
Table of Contents MC-III™ EXP Flow Analyzer Section 3—Configuration and Operation via Keypad.................................................................................. 35 Entering a Calibration Factor . .......................................................................................................................... 36 Entering a Calculated Divisor............................................................................................................................
MC-III™ EXP Flow Analyzer Table of Contents Pulse Output...................................................................................................................................................... 73 Configuring Pulse Output............................................................................................................................ 73 Pulse Output Testing................................................................................................................................
Table of Contents MC-III™ EXP Flow Analyzer Toggle RTS Line Option.............................................................................................................................B-4 Enable Modbus Address Support for Firmware Versions 1.06 through 1.08.............................................B-4 Timing Parameters.....................................................................................................................................B-4 Downloading Options........................
MC-III™ EXP Flow Analyzer Section 1 Section 1—Introduction The NuFlo™ MC-III™ EXP Flow Analyzer (Figure 1.1) packs a full spectrum of gas and liquid measurement functionality, high-speed performance, and log archive and retrieval capabilities in an easy-to-use, explosionproof totalizer. Commonly used operations can be accessed from the six-button keypad on the front of the instrument or from the dynamic interface software, allowing you to calibrate and configure the unit quickly and easily.
Section 1 MC-III™ EXP Flow Analyzer With the press of a single key, totals are saved to nonvolatile memory, minimizing the risk of data loss even if a power outage occurs. Up to 384 daily logs, 768 hourly logs, and 345 event logs can be archived and accessed quickly on demand.
MC-III™ EXP Flow Analyzer Section 1 Table 1.1—MC-III EXP Specifications Electrical Safety Classification Approved by CSA for US and Canada Class I, Div. 1, Groups B, C, D (explosion-proof) Type 4 enclosure T6 temperature class Approved by SIRA to ATEX 10ATEX 1264X IECEx SIR10.0127X II 2 GD E Ex d IIC T6 Gb (-40°C to 70°C) or Ex tD A21 IP66 T85°C (-40°C to 70°C) System Power Internal power supply • 3.
Section 1 MC-III™ EXP Flow Analyzer Table 1.1—MC-III EXP Specifications Inputs Turbine Meter Input Configurable sensitivity adjustment via front panel Sensitivity adjustment range: 20 mV P-P to 40 mV P-P Frequency range: 0 to 3500 Hz Remote Reset Input Optically isolated 3.
MC-III™ EXP Flow Analyzer Section 1 Table 1.
Section 1 MC-III™ EXP Flow Analyzer gallons per minute). Also, the volume unit used for the flow rate can be different from the volume unit used to read Total volume. The unit of measure for the Rate readout and the decimal point position are selected by the operator during calibration. The daily index (Day) display is a two-digit number for selecting a daily archive log for viewing. The number shown here represents the number of days that have passed since the log was saved.
MC-III™ EXP Flow Analyzer Section 1 Enable/disable/set up 4-20 mA output Enable/disable/set up pulse output Enter slave address Enter baud rate Set engineering units for Total readout Set engineering units for Rate readout Set decimal placement for both readouts Calibrate instrument Set flowmeter input sensitivity Select pulse input K-FACTOR MENU OUTPUT MENU DISPLAY MENU Total Day Rate ENTER LOG TEST SAVE RESET TOTAL During calibration, navigates between submenus and scrolls through menu selectio
Section 1 MC-III™ EXP Flow Analyzer Multipoint Linearization The MC-III interface software supports up to 12 calibration points in addition to single-point calibration based on the K-factor provided with the turbine flowmeter. See K-Factor Type, page 66, for more information. Gas Volume Correction Gas turbine meters are calibrated in actual cubic feet (ACF), and measure gas in actual cubic feet.
MC-III™ EXP Flow Analyzer Section 1 Password-protected security access is enabled using the interface software. When this feature is enabled, the user will be prompted for a password when attempting to enter any menu from the keypad. For more information, see Security Setup, page 61.
Section 1 MC-III™ EXP Flow Analyzer Important: The USB communications adapter is approved by CSA for use in hazardous areas. The CSA-approved model is Part No. 2295524-01. The adapter kit, which includes an installation CD, is Part No. 2295634-01. RS-485 IS Barrier (CSA-Approved Units Only) The RS-485 output of a CSA-approved MC-III is approved for intrinsically safe installations.
MC-III™ EXP Flow Analyzer Section 1 • To reset the total with an explosion-proof switch, press and hold the switch for 3 seconds, then release it. See also Explosion-Proof Reset/Control Switch, page 15. • To reset the total with an external pulse generator located in safe area, configure the pulse to be active for 3 seconds. See also Remote Reset Input, page 29. Viewing Daily and Hourly Logs Each day, as the user-defined contract hour passes, a daily flow total is saved to nonvolatile memory.
Section 1 18 MC-III™ EXP Flow Analyzer
MC-III™ EXP Flow Analyzer Section 2 Section 2—Installation The MC-III™ EXP is fully assembled at the time of shipment and ready for mounting to a flowmeter. Field wiring enters the MC-III EXP enclosure through either of two threaded conduit openings in the top of the enclosure. When not in use, the openings are sealed with pipe plugs (Figure 2.1). Ground screws are provided both inside and outside the enclosure.
Section 2 MC-III™ EXP Flow Analyzer • The device has not been assessed as a safety-related device (as referred to by Directive 94/9/EC Annex II, clause 1.5). • Installation shall be carried out by suitably trained personnel in accordance with the applicable code of practice (EN 60079-14 within Europe). • Repair of this equipment shall be carried out by Cameron or in accordance with the instructions in this manual.
MC-III™ EXP Flow Analyzer Section 2 5.00 (127) 5.71 (145) 4.94 (125.5) 8.57 (217.7) approx. Figure 2.2—Direct mount to NuFlo turbine meter with 3/4-in. to 1-in. standoff tube (ATEX) 5.71 (145) 5.71 (145) 4.94 (125.5) 9.29 (236.1) approx. Figure 2.3—Direct mount to a Barton 7000 series flowmeter with 3/4-in. to 3/4-in.
Section 2 MC-III™ EXP Flow Analyzer 5.00 (127) 4.94 (125.5) 7.45 (189.3) 6.56 (16.7) 6.95 (17.65) Figure 2.4—Remote mount on a 2-in. pole using a NuFlo hardware kit (Part No. 9A-30028004) Hazardous Area Installations (CSA) The CSA-approved MC-III EXP has been certified as explosion proof for Class I, Division I, Groups B, C, and D for use in the US and Canada.
MC-III™ EXP Flow Analyzer Section 2 • Replace the lithium battery only with the CSA-approved battery supplied by Cameron (Part No. 9A100005111). • Replace the optional alkaline battery pack only with the CSA-approved battery pack supplied by Cameron (Part No. 9A-50099002). • The lithium battery that powers the MC-III EXP is a sealed unit; however, should a lithium battery develop a leak, toxic fumes could escape upon opening the enclosure.
Section 2 MC-III™ EXP Flow Analyzer 5.71 (145) 5.00 (127) 6.73 (171.1) 1.75 (44.5) Figure 2.5—Direct mount (standard MC-III EXP enclosure) to NuFlo turbine meter with pipe reducer and union (CSA) 5.00 (127.0) 4.55 (115.6) 7.05 (179.1) 1.63 (41.4) Figure 2.
MC-III™ EXP Flow Analyzer Section 2 5.00 (127) 4.94 (125.5) 7.45 (189.3) 6.56 (16.7) 6.95 (17.65) Figure 2.7—Remote mount on a 2-in. pole using a NuFlo hardware kit (Part No. 9A-30028004) Installing the MC-III EXP If the device is to be installed in a hazardous area, configure the device prior to mounting it. See Sections 3 and 4 for instructions. 1. Make sure the flowmeter and magnetic pickup are properly installed as follows: a. Install the turbine flowmeter in the flow line. b.
Section 2 MC-III™ EXP Flow Analyzer Field Wiring The MC-III EXP supports connections for the following: internal power (battery), external power, turbine flowmeter input, pulse input, remote reset input, pulse output, 4 to 20 mA output, flowmeter frequency (amp & square) output, and RS-485 output. Wiring diagrams for each connection are provided below.
MC-III™ EXP Flow Analyzer Section 2 Internal Power Supply The MC-III EXP is shipped with one of three internal power supplies: • a CSA-approved 3.6V lithium battery • an ATEX-approved 3.6V lithium battery • a shrink-wrapped alkaline battery pack containing three C-size industrial-grade batteries (available only with CSA-approved devices) Low-power microprocessor technology enables the MC-III EXP analyzer to operate approximately 2 years on a single lithium battery.
Section 2 MC-III™ EXP Flow Analyzer Turbine Flowmeter (TFM) Input 6-3 0V DC BATTERY The TFM input provides the turbine flowmeter input signal generated by a magnetic pickup, enabling the MCIII EXP to calculate and display instantaneous flow rates and accumulated totals. TB A&S 3 EXT POWER GND J1 TFM B TURBINE MAGNETIC PICKUP VMAX = 3.9 V RESET INPUT A PULSE INPUT TB1 RESET SWITCH J2 Figure 2.
MC-III™ EXP Flow Analyzer Section 2 Remote Reset Input The remote reset input allows the operator to reset the accumulated volume on the MC-III EXP to zero without opening the enclosure. This input is optically isolated. The input is shown connected in two ways, with a power supply and switch in a safe area (Figure 2.11), and with a pulse generator in a safe area (Figure 2.12). The remote reset input can also be controlled with an optional explosion-proof switch mounted in the top of the enclosure.
Section 2 MC-III™ EXP Flow Analyzer Pulse Output The pulse output is a solid-state relay. Each pulse represents a user-defined volume. Because the circuit is isolated, it can be used in conjunction with any other feature on the MC-III EXP. A two-conductor cable from the MC-III EXP to the remote location is required (Figure 2.13). The maximum current rating of the pulse output circuit is 60 mA at 30 VDC. A Zener diode (Part No. 1.5KE56CA) must be installed for CE approval.
MC-III™ EXP Flow Analyzer Section 2 Analog (4-20 mA) Rate Output The 4-20 mA rate output provides a linear current output that represents flow rate. This output requires a twoconductor cable connected to an 8 to 30 VDC power supply (voltage required is dependent on loop resistance) and a current readout device located in the remote location. The 4-20 mA rate output current loop also powers the MC-III EXP.
Section 2 MC-III™ EXP Flow Analyzer Flowmeter Frequency Output The flowmeter frequency (amp & square) output provides an open drain transistor output at the turbine meter frequency, which may be used to provide flow rate and/or total information to peripheral equipment. The output requires a two-conductor cable from the MC-III EXP to the remote frequency readout device requiring 50 mA or less and a 5 to 30 VDC power supply (Figure 2.15).
MC-III™ EXP Flow Analyzer Section 2 RS-485 Output The RS-485 output is required for communication with the interface software. Wiring diagrams are provided for a permanent connection (Figure 2.16), as well as for temporary laptop connections using an RS-485 to RS232 converter (Figure 2.17). The RS-485 output may also be used to install an external communications adapter for connecting a laptop without opening the enclosure.
Section 2 MC-III™ EXP Flow Analyzer Figure 2.
MC-III™ EXP Flow Analyzer Section 3 Section 3—Configuration and Operation via Keypad Calibration of the MC-III™ EXP is a simple matter of entering necessary parameters into the instrument. The process for calibrating the MC-III EXP depends on how the instrument will be used. The keypad can be used to calibrate the MC-III EXP for liquid or gas measurement using preprogrammed units, or for liquid measurement using a calculated divisor.
Section 3 MC-III™ EXP Flow Analyzer Entering a Calibration Factor When the volume is to be expressed in barrels (BBL), gallons (GAL), liters (LIT), cubic meters (M3), or cubic feet (CF), and the flow rate is to be expressed in barrels, gallons, liters, cubic meters or cubic feet per day, per hour, per minute, or per second, the MC-III EXP calculates the divisor automatically; only the turbine meter calibration factor is required.
MC-III™ EXP Flow Analyzer Section 3 Entering a Calculated Divisor When registering the volume in units other than cubic meters, cubic feet, barrels, gallons, or liters, a calculated divisor must be entered in the K-Factor menu instead of the turbine meter calibration factor. Important: When a calculated divisor is used, the units for both the volume display and the volume portion of the flow rate display should be set to USER (no units visible on the display).
Section 3 MC-III™ EXP Flow Analyzer Setting Input Type and Sensitivity The flowmeter signal can be obtained from a magnetic pickup or a pre-amplifier device. The input sensitivity of the MC-III EXP is measured in millivolts (mV) peak-to-peak. This is the threshold value at which the circuitry responds to a signal. If the input signal is less than this value, the MC-III EXP will not count the electrical pulses as a valid turbine meter signal.
MC-III™ EXP Flow Analyzer Section 3 Configuring the Total Display The Total display can be configured for measuring volume in any of five preprogrammed units, any preprogrammed unit times 1,000, or a user-defined unit. Users can specify a decimal point position ranging from 0.1 to 0.0001 of a unit. The device will display volumes up to 10,000,000, however, a maximum of eight digits (including decimals) are visible at a time.
Section 3 MC-III™ EXP Flow Analyzer Configuring the Rate Display The Rate display comprises two parts: a volume unit and a time-base unit. The volume portion of the Rate display can be configured in one of five preprogrammed engineering units or in a user-defined unit (for use with a calculated divisor). To display the volume portion in multiples of 1,000 units, continue scrolling through the volume unit options until the ×1000 entry is displayed, along with the appropriate unit of measure.
MC-III™ EXP Flow Analyzer Section 3 Configuring the 4-20 mA Rate Output Caution: Before performing any 4-20 mA configuration, ensure that all peripheral equipment connected to the 4-20 mA current loop is either disconnected or disabled. Configuring and testing the 4-20 mA output feature on the MC-III EXP with the peripheral equipment in operation may cause false alarms or erroneous operation of the peripheral device or associated equipment.
Section 3 MC-III™ EXP Flow Analyzer To Configure the 4-20 mA Output: Enter the Output menu. Press OUTPUT MENU. Enable or disable the 4-20 mA output. Press UP ARROW to toggle output to “on” or “off”. Press ENTER. If the 4-20 mA output is enabled, the LCD will display the “Low” flow rate and the decimal point in the numeric display will begin blinking. Enter the decimal point position for the “Low” flow rate. The words “4-20 mA LOW” should appear at the bottom of the LCD.
MC-III™ EXP Flow Analyzer Section 3 Configuring the Pulse Output The pulse output feature of the MC-III EXP is typically disabled to reduce current consumption. When the pulse output is not needed, Cameron recommends that this feature be disabled. If the pulse output feature is required, the user will be prompted to enter a pulse output scale factor, which is the volume increment that will cause a pulse output to occur.
Section 3 MC-III™ EXP Flow Analyzer Entering the Slave Address The slave address is a setting used in Modbus® communications. It is a number that ranges from 1 to 65535, excluding 252 to 255 and 64764. If the Modbus® request message contains the matching address, the device will respond to the request. In network arrangements, the device must have a unique slave address. For more information about Modbus® communications, refer to the protocol manual section.
MC-III™ EXP Flow Analyzer Section 4 Section 4—Configuration and Operation via Software The MC-III™ interface software provides easy access to all configuration parameters and is the quickest way to configure the device. While basic parameters can be configured with the keypad (see Section 3), the interface software is the only means of accessing every configurable parameter.
Section 4 MC-III™ EXP Flow Analyzer Connecting the MC-III to a Computer Before attempting to run the interface software, verify that the Scanner is connected to a computer with an RS-485 to RS-232 converter or an RS-485 to USB converter. • See Figure 2.17, page 33, for converter wiring instructions. • See Explosion-Proof Communications Adapter, page A-3, for instructions on installing and wiring an RS485 communications adapter.
MC-III™ EXP Flow Analyzer Section 4 Figure 4.3—Prompt for authorizing time and date synchronization The time difference setting used to generate a synchronize prompt is user-configurable. For more information, see General Options, page B-1. Automating Functions on Software Startup When the software connects with the instrument, the Device Autorun Options screen appears (Figure 4.4). From this screen, users can configure the instrument, download logs, or upload previously saved configuration settings.
Section 4 MC-III™ EXP Flow Analyzer Changing Autorun Settings To change an autorun setting —that is, to initiate an autorun action, to terminate an autorun action, or to change to a different autorun action—without exiting to the Welcome screen, choose Options/Program Options from the menu bar and select Autorun from the dropdown list (Figure 4.5). Figure 4.
MC-III™ EXP Flow Analyzer Section 4 Figure 4.6—Cancel Express Connect option Changing the Communications Port The computer will attempt to connect to the MC-III EXP via the port that the user selects the first time he connects to the instrument. However, if a network dictates that the computer connect through another communications port (actual or virtual), the user must specify the appropriate COM port. To change the COM port setting, click on Options>Select COM Port (Figure 4.
Section 4 MC-III™ EXP Flow Analyzer To select a device from a network of instruments, perform the following: 1. From the Welcome screen, choose File>Cancel Express Connect (Figure 4.6, page 49). 2. Click File>Discover Modbus® Slaves. 3. Enter a range of addresses you want to search (Figure 4.8). 4. Enter the baud rate established for network communications. 5. Adjust the time-out setting, if necessary. 6. Click “Search.” The software will scan all addresses specified and display all instruments connected.
MC-III™ EXP Flow Analyzer Section 4 Setting Log Download Preferences The MC-III EXP automatically saves daily flow logs on the contract hour, and hourly flow logs around the clock. The instrument also automatically creates an event record each time a user change is made. After downloading these logs from the instrument, the instrument saves the data in a file. By default, the logs are saved as .csv files. The file format of the auto-save flow log and event log files is user-configurable.
Section 4 MC-III™ EXP Flow Analyzer Configuring the MC-III EXP The MC-III EXP offers three methods for configuring parameters: • The “Configuration Wizard” (Figure 4.11) condenses the configuration process into nine easy-to-follow steps (compensated gas measurement and multipoint linearization are not supported in the wizard). • The MC-III Main screen (Figure 4.12, page 53) gives users complete access to all configuration parameters. • The Advanced screen (Figure 4.
MC-III™ EXP Flow Analyzer Section 4 Figure 4.12—MC-III Main configuration option for complete access to configurable parameters Figure 4.
Section 4 MC-III™ EXP Flow Analyzer Configuration Wizard The Configuration Wizard is ideal for first-time users who want to perform a basic calibration for liquid or gas measurement. By following the instructions on nine screens, even a first-time user can successfully configure the MC-III EXP. See Table 4.1, page 55, for a list of settings that are configurable with the keypad, the software Configuration Wizard, and the software MC-III Main Screen.
MC-III™ EXP Flow Analyzer Section 4 Table 4.
Section 4 MC-III™ EXP Flow Analyzer MC-III Main Screen The MC-III Main menu screen (Figure 4.15) is the configuration hub of the MC-III EXP. From the Main screen, users can • review all current instrument settings and flow readings • check the log archive and instrument status • access all submenus for changing configurable parameters • reset the total to zero Figure 4.15—MC-III Main screen provides direct access to all menu functions using the scrolling menu bar at the left side of the screen.
MC-III™ EXP Flow Analyzer Section 4 Buttons and Tools Apply and OK Buttons Changing parameters on a submenu screen involves selecting the proper screen from the scroll bar, selecting information from dropdown menus or entering data in data fields, and saving the data using the “Apply” or “OK” button. “Apply” allows the user to apply the changes without leaving the submenu screen. “OK” allows the user to apply the changes and returns the user to the Main screen.
Section 4 MC-III™ EXP Flow Analyzer Configuration Submenus (See Scroll Bar on MC-III Main Screen) System Setup Wellsite Information K-Factor Entry 58 Communications Port Turbine Input 4-20 mA Output
MC-III™ EXP Flow Analyzer Section 4 Configuration Submenus (cont’d) Pulse Output Flow Archive Event Archive 59
Section 4 MC-III™ EXP Flow Analyzer System Setup The first of the submenus on the Main screen scroll bar—System Setup—allows users to adjust time/date, set the desired contract hour, enable or disable the password-protected security option, and adjust the LCD contrast (Figure 4.17). Figure 4.17—System Setup screen Time/Date Synchronization The MC-III EXP includes a real-time clock for timekeeping and log time stamps.
MC-III™ EXP Flow Analyzer Section 4 Security Setup Setting a keypad security code will prevent unauthorized personnel from altering calibration data or resetting totals, and the security function is recommended to preserve data integrity of the system. The MC-III EXP’s keypad security feature is disabled at the factory. To enable the security protection, check the “Keypad Lock” checkbox on the System Setup screen, and enter a four-digit lock code in the field below the checkbox. (Do not use “0000.
Section 4 MC-III™ EXP Flow Analyzer To change the slave address, select a number using the “plus” and “minus” buttons on the screen, or using the page-up and page-down keys on your computer keyboard. If Modbus® communication is not used, leave the slave address at the factory setting (1). Baud Rate The baud rate is the number of bits per second that are on the serial port. Baud rates supported by the MC-III EXP range from 300 to 115.2K.
MC-III™ EXP Flow Analyzer Section 4 Bus Delay Bus delay is the amount of time (in milliseconds) that passes before the MC-III EXP attempts to take control of the RS-485 bus and transmit a message back to the requesting device. The MC-III EXP responds very quickly to incoming Modbus® requests—in some cases, too quickly. A configurable delay allows the user to control this response time.
Section 4 MC-III™ EXP Flow Analyzer Turbine Input The Turbine Input screen (Figure 4.21) allows users to configure the displays for volume and rate, select the type of input to be used (turbine input or pulse input), determine cut-off thresholds for measuring flow, and determine the display update frequency. Figure 4.
MC-III™ EXP Flow Analyzer Section 4 Rate Display The Rate display comprises two parts: a volume unit and a time-base unit. The volume portion of the Rate display can be configured in one of five preprogrammed engineering units or in a user-defined unit (for use with a calculated divisor). To display the volume portion in multiples of 1,000 units, check the “Multiplier” checkbox.
Section 4 MC-III™ EXP Flow Analyzer K-Factor Entry The K-Factor Entry screen (Figure 4.23) allows users to calibrate the MC-III EXP using a single calibration factor from a turbine flowmeter or multipoint linearization. Users can also configure the MC-III EXP to compensate for the effect of pressure, temperature, and compressibility on gas volume measurements. Figure 4.23—K-Factor Entry screen K-Factor Units K-Factor entry consists of selecting the K-Factor unit and type (linear or multipoint).
MC-III™ EXP Flow Analyzer Section 4 For multipoint calibration, the user selects the Multipoint checkbox, and then enters the number of calibration points he desires using the “plus” and “minus” buttons on the screen or the page-up and page-down keys on a computer keyboard. Up to 12 calibration points may be entered. Frequency and factor data-entry fields are automatically activated for the number of calibration points selected. The user can then enter the appropriate data for each point.
Section 4 MC-III™ EXP Flow Analyzer Gas Volume Correction (Supercompressibility Calculation) Gas turbine meters are calibrated in actual cubic feet (ACF), and measure gas in actual cubic feet. In some applications, a user may benefit from referencing gas measurements back to standard conditions by measuring in terms of standard cubic feet (SCF). The MC-III EXP makes this process quick and easy, by using fixed average values for the flowing gas temperature and flowing gas pressure.
MC-III™ EXP Flow Analyzer Section 4 6. Enter a known compressibility factor, or press “Calculate compressibility from gas comp.” to view a selection of gas compositions (Figure 4.27). • • To automatically calculate the compressibility factor, enter the gas composition and click on “Calculate.” The software will calculate flowing gas compressibility in accordance with the AGA-8 detail method. The K-Factor Entry screen will reappear, and the calculated super-compressibility factor will be displayed.
Section 4 MC-III™ EXP Flow Analyzer 4-20 mA Output Caution: Before performing any 4-20 mA calibration, ensure that all peripheral equipment connected to the 4-20 mA current loop is either disconnected or disabled. Calibrating and testing the 4-20 mA output feature on the MC-III EXP with the peripheral equipment in operation may cause false alarms or erroneous operation of the peripheral device or associated equipment. The 4-20 mA Output screen (Figure 4.
MC-III™ EXP Flow Analyzer Section 4 IMIN = minimum current output (4 mA) High = programmed flow rate that produces a 20-mA output Low = programmed flow rate that produces a 4-mA output RATECURR= flow rate Figure 4.29—4-20 mA output configuration Enabling 4-20 mA Output By default, the 4-20 mA output option is disabled. To enable this feature, perform the following steps: 1. Check the “Enable” checkbox. 2. Click on the pencil icon next to the Low Flow Rate field to launch a data-entry window (Figure 4.
Section 4 MC-III™ EXP Flow Analyzer To use the test mode, press the “Test 4-20 mA Output” button at the bottom of the 4-20 mA Output screen. The test page will appear (Figure 4.31), and the 4-20 mA Output option will be temporarily disabled. To test the output, enter a value in the Output Current (mA) field, and press “Set Output.” Figure 4.
MC-III™ EXP Flow Analyzer Section 4 Pulse Output The Pulse Output screen (Figure 4.32) allows users to configure the MC-III EXP to provide a pulse output representing increments in volume. A test mode function is also included, allowing a user to calibrate and/or verify the output received by an end device. Figure 4.32—Pulse Output screen Configuring Pulse Output By default, the pulse output option is disabled to reduce current consumption.
Section 4 MC-III™ EXP Flow Analyzer Output screen. In applications where short bursts of high flow rates are common (separators, tank loading, dump cycles, etc.), the displayed flow rate can exceed the maximum flow rate for short periods without compromising the accuracy of the total represented by the pulse output. Figure 4.33—Data-entry screen for pulse output scale factor Pulse Output Testing The MC-III software allows users to test the pulse output by writing a value to the pulse accumulator.
MC-III™ EXP Flow Analyzer Section 4 Saving and Uploading Configuration Files The MC-III EXP allows a user to save configuration settings in a file that is stored on the user’s computer and can be uploaded to the MC-III EXP as needed. Configuration files are easily identified by the .mc3 extension. The default filename begins with “config” and uses the following structure: File Format .mc3 Filename Structure “config”
Section 4 MC-III™ EXP Flow Analyzer 2. Click “Save.” An “Information” window will appear to confirm that the file has been saved. The “wellname” folder will bear the well name that appears on the Wellsite Information screen. If no well name is entered on this screen, the folder will be named “NO_WELLNAME” (Figure 4.36, page 75). Uploading a Configuration File To upload a configuration file, perform the following steps: 1. From the Main screen, select File>Return to Welcome Screen (Figure 4.37). 2.
MC-III™ EXP Flow Analyzer Section 4 Figure 4.38—Upload a Configuration option Figure 4.
Section 4 MC-III™ EXP Flow Analyzer Advanced Access System configurators and host programmers can also access the software through the Advanced Access menu (Figure 4.40). This option provides access to the Modbus® registers for the MC-III EXP. A series of tabs on the right side of the screen (real time, input configuration, output configuration...) allows the user to view a specific group of registers, or to view all registers in one list.
MC-III™ EXP Flow Analyzer Section 5 Section 5—Flow Logs and Event Logs The MC-III™ EXP’s flow archive expands the user’s ability to track flow volume over time by allowing the user to view flow data in tabular and trend formats, to save or print log data in reports, and to export log data into a spreadsheet. Additionally, an event archive allows users to identify changes to configuration parameters and other settings for troubleshooting operational issues.
Section 5 MC-III™ EXP Flow Analyzer Figure 5.2—Log download options Log Directory and Filenames Auto-save log files are saved in C:\NuFlo log data\MC-III\. The “wellname” folder will bear the well name that appears on the Wellsite Information screen. If no well name is entered on this screen, the folder will be named “NO_WELLNAME.” Log filenames incorporate the serial number of the MC-III EXP, the date, the time, and a capital letter indicating the log type, as shown below. File Format .
MC-III™ EXP Flow Analyzer Section 5 Figure 5.3—Flow Archive screen Downloading Flow Logs To download flow logs, click the “Download” button on the Flow Archive screen, and select either Download All Flow Logs or Download Only New Flow Logs (Figure 5.4). (Download Only New Flow Logs will display only flow logs that have been created since the last download was performed.) Tabular views of the archive record (Figure 5.
Section 5 MC-III™ EXP Flow Analyzer Figure 5.5—Downloaded flow logs (tabular view) Viewing Trend Charts In the daily and hourly trend views, flow volumes are charted on a grid, with flow record numbers forming the horizontal axis, and flow volumes shown on the vertical axis. A graphical line defined by two yellow endpoints represents the range of archived logs available for viewing. A yellow triangle marks the first record, and a yellow circle marks the final record.
MC-III™ EXP Flow Analyzer Section 5 Figure 5.6—Flow Archive screen, trend view Printing/Saving a Report A flow log can be printed and/or saved as a table (Figure 5.7) or a trend chart (Figure 5.8, page 84). To print a log, select the appropriate tab at the bottom of the Flow Archive screen, and press the “Print” button in the upper right corner of the Flow Archive screen. (If the display is tabular, the button will read “Print Table”; if the display is a chart, the button will read “Print Chart.
Section 5 MC-III™ EXP Flow Analyzer Figure 5.8—Print preview, trend view To save the report, perform the following steps: 1. Select File>Save from the task bar (Figure 5.8). A “Save Report As” window will open (Figure 5.9). 2. Enter a filename. The filename extension will be .rps. 3. Click “Save.” By default, log archive reports are saved in C:\NuFlo log data\MC-III\. However, the user can specify a new location, if desired.
MC-III™ EXP Flow Analyzer Section 5 Viewing a Saved Report To view a saved report, perform the following steps: 1. From the Flow Archive screen, click “Print Table” (or “Print Chart” for trend chart views) in the upper right corner of the screen. 2. Select File>Load from the task bar at the top of the page (Figure 5.10). A Load Report window will open (Figure 5.11). 3. Click on the report you wish to view. A preview of the report will appear in the right portion of the window. 4. Click “Open.” Figure 5.
Section 5 MC-III™ EXP Flow Analyzer Exporting Flow Logs Flow logs can be directly exported to an .xls or .csv file for ease in viewing and distributing. To export a file, perform the following steps: 1. Click the “Export Data” button near the top of the Flow Archive screen and select the .xls or .csv format (Figure 5.12). 2. When the “Export Daily (or Hourly) Log Data...” window appears, click “Save.
MC-III™ EXP Flow Analyzer Section 5 Figure 5.13—Default directory for exported log files Event Archive The Event Archive screen (Figure 5.14) allows users to download, view, export, and print up to 345 user event logs. Event logs are generated to track user changes such as K-Factor changes, input setting changes, poweron and “watch-dog” resets, and the flow and frequency cut-off settings. Figure 5.
Section 5 Downloading Event Logs MC-III™ EXP Flow Analyzer To download event logs, click the “Download” button on the Event Archive screen, and select either Download All Event Logs or Download Only New Event Logs, (Figure 5.15). Download Only New Event Logs will display only flow logs that have been created since the last download was performed. Tabular views of the archive record (Figure 5.
MC-III™ EXP Flow Analyzer Section 5 Printing/Saving a Report To print an event log, press the “Print Table” button in the upper right corner of the Event Archive screen. A Print Preview screen will appear, displaying the image to be printed. To print the report, select File>Print from the task bar. To save the report, perform the following steps: 1. Select File>Save from the task bar. A “Save Report As” window will open. 2. Enter a filename. The filename extension will be .rps. 3. Click “Save.
Section 5 Figure 5.
MC-III™ EXP Flow Analyzer Section 6 Section 6 - MC-III EXP Maintenance The MC-III™ EXP is designed to provide many years of service with minimal maintenance. Batteries require periodic replacement, and battery life depends on whether battery power is the primary or secondary power source. All configuration settings are stored in nonvolatile memory; therefore, configuration settings will not be lost in the event of battery failure.
Section 6 MC-III™ EXP Flow Analyzer Remove screws to access battery compartment Figure 6.1— Removal of faceplate/keypad assembly 3. Lift the display/keypad assembly from the enclosure, making sure the circuit assembly does not contact the enclosure. 4. Loosen the velcro strap, disconnect the battery from the J1 connector on the circuit assembly, and remove the battery from the enclosure (Figure 6.2). Figure 6.
MC-III™ EXP Flow Analyzer Section 6 5. Install the new battery in the enclosure in the same position as the original battery, and secure the Velcro tightly around the battery. 6. Connect the replacement battery to the J1 connector. 7. Place the circuit assembly over the standoffs and fasten with the two #4-40 × 7/8” screws, ensuring that all connector wiring is inside the enclosure. 8. Replace the enclosure cover, threading it onto the enclosure in a clockwise direction.
Section 6 MC-III™ EXP Flow Analyzer Remove screws to access battery compartment Figure 6.3—Removal of faceplate/keypad assembly 3. Lift the display/keypad assembly from the enclosure. 4. Loosen the Velcro strap, disconnect it from the J1 connector on the circuit assembly, and remove the battery from the enclosure (Figure 6.4). Figure 6.
MC-III™ EXP Flow Analyzer Section 6 5. Install the new shrink-wrapped battery pack in the same position as the original battery pack as shown in Figure 6.4, page 94 and secure the Velcro strap to hold the pack tightly in place. 6. Connect the replacement battery pack to the J1 connector. 7. Place the circuit assembly over the standoffs and fasten with the two #4-40 × 7/8” screws, ensuring that all connector wiring is inside the enclosure. 8.
Section 6 MC-III™ EXP Flow Analyzer Figure 6.5—Disassembly of circuit board/keypad assembly 8. Disconnect the keypad ribbon cable from the J3 connector on the LCD side of the circuit assembly as follows: a. Grasp the black clip between a thumb and forefinger (see white arrows in Figure 6.6 below). b. Squeeze both sides of the clip and gently pull to release the clip from the plastic connector that holds it in place. DO NOT PULL on the ribbon cable.
MC-III™ EXP Flow Analyzer Section 6 9. Remove the original circuit assembly from the enclosure. 10. Connect the ribbon cable of the keypad to the J3 connector on the LCD side of the circuit assembly as follows: a. Insert the end of the ribbon cable into the plastic clip. b. While holding the ribbon cable in place, press the black plastic clip into the connector until it snaps. 11. Mount the circuit assembly to the keypad with the two #4-40 × 5/16” screws. 12.
Section 6 MC-III™ EXP Flow Analyzer 9. Mount the display/keypad assembly to the enclosure with the two #4-40 × 7/8” screws. 10. Recalibrate the MC-III EXP if necessary and replace the enclosure cover. Firmware Update Before installing new firmware or moving the instrument to a new site, it is recommended that the user clear the EEPROM, which will erase all configuration settings and log data.
MC-III™ EXP Flow Analyzer Section 6 MC-III Parts List (CSA) CAUTION: EXPLOSION HAZARD—SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR CLASS I, DIVISION 1 IN THE UNITED STATES AND CANADA. USE OF SPARE PARTS OTHER THAN THOSE IDENTIFIED BY CAMERON VOIDS CSA CERTIFICATION. CAMERON BEARS NO LEGAL RESPONSIBILITY FOR THE PERFORMANCE OF A PRODUCT THAT HAS BEEN SERVICED OR REPAIRED WITH PARTS THAT ARE NOT AUTHORIZED BY CAMERON.
Section 6 MC-III™ EXP Flow Analyzer MC-III Parts List (ATEX) CAUTION: EXPLOSION HAZARD—SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR ATEX AND IECEX APPROVALS. USE OF SPARE PARTS OTHER THAN THOSE IDENTIFIED BY CAMERON VOIDS ATEX/IECEX CERTIFICATION. CAMERON BEARS NO LEGAL RESPONSIBILITY FOR THE PERFORMANCE OF A PRODUCT THAT HAS BEEN SERVICED OR REPAIRED WITH PARTS THAT ARE NOT AUTHORIZED BY CAMERON.
MC-III™ EXP Flow Analyzer Appendix A Appendix A—MC-III EXP Hardware Options Explosion-Proof Reset/Control Switch An external explosion-proof reset/control switch (Figure A.1) allows the user to zero the total and manually select which parameter is displayed on the LCD and view daily logs instantaneously without removing the instrument cover or connecting the instrument to a laptop computer. Caution: This switch is available in two models.
A&S TB1 3 EXT POWER GND TB J1 6-3 0V DC MC-III™ EXP Flow Analyzer BATTERY Appendix A TB2 RESET INPUT RS485 PULSE SLAVE OUT PULSE INPUT TFM 4-20 OUT RESET SWITCH J2 Remote switch connector (Part No. 9A-50053008) Figure A.2—Explosion-proof switch wiring Operation Zeroing the Total. To reset the volume total with an explosion-proof switch, press and hold the switch for at least 3 seconds, then release it.
MC-III™ EXP Flow Analyzer Appendix A Additionally, when the log viewing function times out, the accumulated total is automatically saved to memory. Saving the Total. When the daily log display times out, the MC-III EXP will automatically save the accumulated total. Explosion-Proof Communications Adapter The explosion-proof communications adapter provides an RS-485 connection for connecting a laptop or PC to the instrument without removing the instrument cover.
Appendix A MC-III™ EXP Flow Analyzer COM ADAPTER SOCKET PLUG CONNECTOR (LEFT AND RIGHT WIRE POSITIONS ARE VALID ONLY WHEN PLUG IS FACING SOCKET IN POSITION SHOWN) TD(A) TD(B) GND RS-485 CONVERTER RS-232 9-PIN CONNECTOR GND +12V Part No. 101283116 Figure A.5—Wiring of plug connector to 9-pin RS-232 to RS-485 converter cable, Part No. 9A-101283116 Connecting to the Communications Adapter Take the following precautions when disassembling the communications adapter.
MC-III™ EXP Flow Analyzer Appendix A Disconnecting the Communications Adapter To disconnect the adapter, perform the following steps: 1. Remove the plug connector (with converter cable attached) from the socket. 2. Place the blanking plug inside the union nut (removed in step 1). 3. Screw the union nut onto the union half to cover the socket. Hand-tighten to ensure a snug connection. Important: Do not disconnect the plug connector from the RS-232 to RS-485 converter cable when it is not in use.
Appendix A MC-III™ EXP Flow Analyzer USB Communications Adapter The NuFlo™ USB Adapter (Figure A.7) allows a user to connect a computer to the MC-III using a standard off-the-shelf USB connector cable for quick and easy downloads without opening the MC-III enclosure. Caution: This adapter is CSA-approved for use in hazardous areas. It does not have ATEX/IECEX approval. The USB adapter is comprised of a USB adapter socket, a blanking plug, and a union nut (Figure A.8).
MC-III™ EXP Flow Analyzer Appendix A User-Supplied Hardware To connect the USB adapter to an MC-III using MC-III software, a user-supplied universal serial bus USB A/B cable is required (Figure A.9). Figure A.9—User-supplied USB A/B cable Using the Adapter The MC-III software CD supplied with the NuFlo USB Adapter contains the drivers required to enable USB communications for an MC-III when the NuFlo USB Adapter is installed.
Appendix A MC-III™ EXP Flow Analyzer 3. Connect the adapter cable to the RS-485 port on the MC-III EXP main circuit board (Figure A.6, page A-5). 4. Connect the plug connector to an RS-485 converter cable (Figure A.5, page A-4), if applicable (black wire to negative terminal). RS-485 SLAVE USB CONNECTOR 4-20 OUT + – + – PULSE OUT TB2 Figure A.10—Wiring of NuFlo USB adapter (required only when purchased as a kit) 5.
MC-III™ EXP Flow Analyzer Appendix B Appendix B—Software Program Options In addition to the standard configuration and flow log menus, the MC-III™ interface software includes a Program Options menu that allows users to customize the way the software functions, handles log data, and communicates. To access the Program Options menu, select Options in the menu bar; Program Options is the last entry in the dropdown list (Figure B.1). Figure B.
Appendix B MC-III™ EXP Flow Analyzer Figure B.2—Threshold setting for internal clock synchronization Autorun Options When the software connects with the MC-III EXP, the Device Autorun Options screen appears, prompting the user to select any of six actions, including navigating to one of three configuration screens, downloading flow and event logs, and uploading a configuration file.
MC-III™ EXP Flow Analyzer Appendix B Figure B.3—Menu for changing autorun option selection Communications Options The Communications Options screen (Figure B.4) allows users to customize communication parameters for optimizing performance, depending on the communication path. Figure B.
Appendix B MC-III™ EXP Flow Analyzer Express Connect Option By default, the MC-III interface software automatically attempts to connect to the instrument when a user clicks on the desktop application icon. This feature is controlled by the “Always attempt Express Connect on program startup” checkbox setting in the Communications Option screen (Figure B.4, page B-3).
MC-III™ EXP Flow Analyzer Appendix B the device on, and then waits the specified length of time before attempting to transmit data. By default, this parameter is set to zero. Number of Retries By default, the computer will make two attempts to connect to the MC-III EXP before providing an error message.
Appendix B MC-III™ EXP Flow Analyzer To save the logs in .xls format, check the “XLS-MS Excel File” checkbox (Figure B.5, page B-5). Additional checkboxes allow for changes in the appearance of downloaded log data. Users can also export flow data to a .csv or .xls file as needed. For complete information, see Exporting Flow Logs, page 86. Advanced Options Clear EEPROM The Clear EEPROM function will erase all configuration settings and log data.
MC-III™ EXP Flow Analyzer Appendix C Appendix C—Lithium Battery Information Lithium Battery Disposal Once a lithium battery is removed from a device and/or is destined for disposal, it is classified as solid waste under EPA guidelines. Depleted lithium batteries are also considered to be hazardous waste because they meet the definition of Reactivity, as per 40 CFR 261.23(a)(2), (3) and (5).
Appendix C C-2 MC-III™ EXP Flow Analyzer
MC-III™ EXP Flow Analyzer Appendix D Appendix D—Communications Protocol Firmware Version 2.03 Register Table Version 1 Introduction The communications protocol for the MC-III is in accordance with Modicon, Inc. RTU Mode Modbus® as described in Modicon Modbus Protocol Reference Guide, PI-MBUS-300 Rev. J, June 1996. All registers are implemented as 4X or holding registers. Reading of registers is implemented via function code 03H (Read Holding Registers).
Appendix D MC-III™ EXP Flow Analyzer The Floating Point-32 bit (FP32) data type follows the IEEE-754 format and consists of 32 bits contained in a single register. The standard word ordering for multiple register data types, such as floating-point numbers or long integers, is for the most significant word to appear first in the message. For master devices requiring the least significant word to appear last, a subset of registers is provided on page D-11.
MC-III™ EXP Flow Analyzer Appendix D Note All registers cited in this document refer to the address of the register that appears in the actual Modbus® message. For example, register 17001 has an address of 0x4268 Hexadecimal in the message. Note Two sets of 16-bit holding registers are provided (starting at 17001 and 8000). The “8000” set should be used with master devices that require the holding registers to be in the range of 40000 to 49999.
Appendix D MC-III™ EXP Flow Analyzer System Configuration Register (Decimal) Register (Hex) 1016 03F8 Description LCD Contrast [0-31] Data Type Access Default U16 R/W 6 Product Code (register 1000) The Product Code is a read-only parameter used for identification. This parameter is set at the factory and it will always read 0x40 hexadecimal (64 decimal).
MC-III™ EXP Flow Analyzer Appendix D Real Time Register (Decimal) Register (Hex) Data Type Access 1200 04B0 Year (Real Year = register value plus 2000) U16 R/W 1201 04B1 Month [1-12] U16 R/W 1202 04B2 Day [1-31] U16 R/W 1203 04B3 Hour [0-23] U16 R/W 1204 04B4 Minute [0-59] U16 R/W 1205 04B5 Second [0-59] U16 R/W Description This block of registers is used to set the instrument’s internal clock.
Appendix D MC-III™ EXP Flow Analyzer Input Configuration Register (Decimal) Register (Hex) Data Type Access Default U16 R/W 2 07D3 Rate Time Base 0 = Second 1 = Minute 2 = Hour 3 = Day U16 R/W 3 2004 07D4 Rate Decimal Point Location 0 = No digits to the right of the decimal point 1 = Tenths 2 = Hundredths 3 = Thousandths U16 R/W 2 2005 07D5 Calculation Period (sec) U16 R/W 4 2006 07D6 Turbine Meter Input Configuration 0 = Pulse Input 1 = TFM - LOW 2 = TFM - MED 3 = TFM - HIGH U
MC-III™ EXP Flow Analyzer Appendix D Input Configuration Register (Decimal) Data Type Access Default 07D9 Factor Units 0 = No Units 1 = Gallons 2 = Liters 3 = Barrels 4 = Cubic Feet 5 = Cubic Meters U16 R/W 1 2010 07DA Factor Decimal Point Location 0 = No digits to the right of the decimal point 1 = Tenths 2 = Hundredths 3 = Thousandths 4 = Ten Thousandths U16 R/W 2 2011 07DB Calibration Frequency [1] (Hz)* FP R/W 1.00 2013 07DD Calibration Frequency [2] (Hz)* FP R/W 1.
Appendix D MC-III™ EXP Flow Analyzer Input Configuration Register (Decimal) Register (Hex) Data Type Access Default 2051 0803 Calibration Factor [9]* (pulses per configured factor unit) FP R/W 1.00 2053 0805 Calibration Factor [10]* (pulses per configured factor unit) FP R/W 1.00 2055 0807 Calibration Factor [11]* (pulses per configured factor unit) FP R/W 1.00 2057 0809 Calibration Factor [12]* (pulses per configured factor unit) FP R/W 1.
MC-III™ EXP Flow Analyzer Appendix D Holding Registers (16-bit Mode) Register (Decimal) Register (Hex) Data Type Access 17001 (8000) 4269 (1F40) Hourly Pointer [1 to 768] FP RO 17003 (8002) 426B (1F42) Daily Pointer [1 to 384] FP RO 17005 (8004) 426D (1F44) Event Counter [1 to 345] FP RO 17007 (8006) 426F (1F46) Real Date (MM/DD/YY) FP RO 17009 (8008) 4271 (1F48) Real Time (HH:MM:SS) FP RO 17011 (8010) 4273 (1F4A) Grand Total (in terms of configured volume units) FP RO 17
Appendix D MC-III™ EXP Flow Analyzer Holding Registers (16-bit Mode) D-10 Register (Decimal) Register (Hex) Data Type Access 17043 (8042) 4293 (1F6A) Supply Voltage (VDC) FP RO 17045 (8044) 4295 (1F6C) Battery Voltage (VDC) FP RO 17047 (8046) 4297 (1F6E) Grand Total (base unit) FP RO 17049 (8048) 4299 (1F70) Instantaneous Flow Rate (base unit) FP RO 17051 (8050) 429B (1F72) Daily Total (base unit) FP RO 17053 (8052) 429D (1F74) Daily Run Time (seconds) FP RO 17055 (8054
MC-III™ EXP Flow Analyzer Note: Appendix D The registers below should be used with master devices that require the floating point numbers to be presented with the least significant word first. Their counterparts with floating point numbers in reverse word order—registers 17011, 17013, 17027 and 17015—are listed in the previous table.
Appendix D MC-III™ EXP Flow Analyzer Polling Registers The MC-III EXP device stores the volume accumulated since the last polling sequence in a set of polling registers (17023). In addition, the instrument monitors the number of seconds of flow time in each polling interval (17025). There is also an index to the number of polls requested (17039). The polling sequence is started by writing to the Control Register.
MC-III™ EXP Flow Analyzer Appendix D The unit of measurement for this register is pulses per configured factor unit. Once the K-factor is calculated based on frequency and K-factor points, this value is multiplied by 1000 if the user chooses to read the volume in units x 1000, and it is divided by the Standard Measurement Correction Factor (register 2059).
Appendix D MC-III™ EXP Flow Analyzer Holding Registers (32-bit Mode) D-14 Register (Decimal) Register (Hex) Description Data Type Access 7019 1B6B Polling Index [0-65535] FP32 RO 7020 1B6C Internal Temp (°C) FP32 RO 7021 1B6D Supply Voltage (VDC) FP32 RO 7022 1B6E Battery Voltage (VDC) FP32 RO 7023 1B6F Grand Total (base unit) FP32 RO 7024 1B70 Instantaneous Flow Rate (base unit) FP32 RO 7025 1B71 Daily Total (base unit) FP32 RO 7026 1B72 Daily Run Time (seconds
MC-III™ EXP Flow Analyzer Appendix D Control Register Register (Decimal) Register (Hex) 70 0046 Description Control Register #1 Data Type Access U16 R/W The Control register allows specific functions to be implemented via the communication port. The following table shows the value to be written to the control register to implement the desired function.
Appendix D MC-III™ EXP Flow Analyzer The following registers are used for hourly, daily and event log registers. Hourly and daily records comprise five values (registers). See the Enron Hourly/Daily Record Format, page D-16. Note that event records are downloaded one at a time.
MC-III™ EXP Flow Analyzer Appendix D Enron Event Record Format Parameter Data Type Status U16 Address U16 Time (HH:MM:SS) FP32 Date (MMDDYY) FP32 As-Found FP32 As-Left FP32 The Event Log status is a 16-bit value used to indicate the reset source in the event log. The following table indicates the reset status.
Appendix D D-18 MC-III™ EXP Flow Analyzer
WARRANTY - LIMITATION OF LIABILITY: Seller warrants only title to the products, software, supplies and materials and that, except as to software, the same are free from defects in workmanship and materials for a period of one (1) year from the date of delivery. Seller does not warranty that software is free from error or that software will run in an uninterrupted fashion. Seller provides all software “as is”.
+603.5569.0501 ms-kl @ c-a-m.