Specifications
10 4114V104-UK
Order codes
4114 = Universal transmitter 
4501 = Display / programming front 
5910 = CJC connector
Electrical specifications
Specifications range ....................................  -20°C to +60°C
Common specifications:
Supply voltage, universal ............................ 21.6...253 VAC, 50...60 Hz or
  19.2...300 VDC
Max. consumption.......................................  ≤ 2.0 W 
Fuse .............................................................  400 mA SB / 250 VAC 
Isolation voltage, test / operation ............... 2.3 kVAC / 250 VAC 
Communications interface .......................... Programming front 4501 
Signal / noise ratio ...................................... Min. 60 dB (0...100 kHz) 
Response time (0...90%, 100...10%): 
  Temperature input ....................................  ≤ 1 s 
  mA / V input .............................................  ≤ 400 ms 
Calibration temperature............................... 20...28°C 
Accuracy, the greater of the general and basic values:
General values
Input 
type
Absolute 
accuracy
Temperature 
coefficient
All ≤ ±0.1% of span ≤ ±0.01% of span / °C 