SM425 INSTRUCTION MANUAL    5.0 WIRING 
Figure 5-9 Power and Grounding wiring 
5.5.1 Power Wiring 
Connect the power wiring to the L and N terminals of the converter. Proceed as follows: 
1.  Use a 2-core or a 3-core shielded cable. 
2.  The size of converter terminal screw threads is M4. Each cable should be terminated with the 
corresponding crimp-on terminal. 
5.5.2 Grounding Wiring 
The ground wiring of the detector should be connected to the ground terminal of the detector case. The 
ground wiring of the converter should be connected to either the ground terminal of the converter case or 
the protective ground terminal in the equipment. The grounding terminals of the detector and the 
converter are of size M4. Proceed as follows: 
1.  Keep ground resistance to 100Ω or less. 
2.  If the ambient temperature of the wiring installation is 80 to 150°C for the detector wiring, use 
wiring material with sufficient heat resistance. 
3.  When connecting the ground wiring to the ground terminal of the converter case, be sure the lock 
washer is in contact with the case surface (see Figure 5-9.). 
4.  Ensure the jumper plate is connected between the “G” terminal and the protective ground 
terminal of the converter. 
5.6 Contact Output Wiring 
Contact outputs 1 to 3 can be user programmed to "low limit alarm", "high limit alarm", etc. The function of 
contact output 4 is fixed ("error output") and its action (contact closed during error condition) cannot be 
changed. 
When using these contact outputs, install the wiring as follows: 