Instruction manual
8.0 DETAILED DATA SETTING    SM425 INSTRUCTION MANUAL 
8.2 Configuring Output Hold 
The “output hold” function retains an analog output signal at a predetermined value during certain 
equipment states. Outputs 1 and 2 can be set independently. Table 8-2 shows the output hold selections 
available for each of the states. 
Table 8-2 Output Hold Selections 
Equipment states  During warm-
up 
Under 
maintenance 
Under calibration 
During blow back 
On Error 
Occurrence 
(*1) 
Output hold 
selections 
available 
4 mA 
○ 
20 mA 
○ 
No hold feature   
○  ○  ○ 
Retains output 
from just before 
occurrence 
○  ○  ○ 
Preset value 
(2.4 to 21.6 mA) 
○  ○  ○  ○ 
○: The output hold functions are available 
*1: The output hold functions for on error occurrence are unavailable when option code “/C2” or “C3” (NAMER NE 43 compliant) is 
specified. 
8.2.1 Definition of Equipment States 
1.  During warm up: “During warm up” is the time required after applying power until the sensor 
temperature stabilizes at 750°C, and the equipment is in the measurement mode. 
2.  Under maintenance: “Under maintenance” is the time starting when the Setup key on the Basic 
Panel display screen is pressed until returning back to the Basic Panel display screen. The sub-
screens under the Execution/Setup display screen are included. See Figure 7-4. 
3.  Under calibration (see Section 0, Calibration): For manual calibration “Under calibration” is the 
time starting from entering the Manual calibration display screen (shown in Figure 8-1) until the 
Calibration End key is pressed and the preset output stabilization time elapses. 
Figure 8-1 Manual calibration 
42500001 REV A    121 










