Instruction manual
8.0 DETAILED DATA SETTING    SM425 INSTRUCTION MANUAL 
Figure 8-3 mA-outputs preset Selection 
2.  Select the desired equipment state. Figure 8-3 shows an example of selecting Maintenance. 
Select the desired output hold selection. 
3.  If a preset value is selected, set the corresponding output current. Select the preset value just 
below Maintenance on the screen; the numeric-data entry display screen appears. Enter the 
desired current value. To set 10 mA, type in 100 and press the [Enter] key to complete the 
setting. The setting range is from 2.4 to 21.6 mA. 
NOTE: The "Error" selection on the “mA-outputs presets” display screen is not available for option codes 
"/C2" or "/C3" (NAMUR NE 43 compliant). 
8.2.4 Default Output Hold Settings 
Default output hold settings are shown in Table 8-3. 
Table 8-3 Output Hold Default Values 
8.3 Configuring Oxygen Concentration Alarms 
The SM423 has four programmable alarms based on oxygen concentration: high-high, high, low, and low-
low. This paragraph explains the operation and programming of these alarms. 
8.3.1 Alarm Values 
1.  High-high and high alarm values: These alarms occur if oxygen levels exceed the alarm set 
points and the alarms are enabled (set to ON). The range for these set points is 0 to 100% O2. 
2.  Low and low-low alarm values: These alarms occur if oxygen levels fall below the alarm set 
points and the alarms are enabled (set to ON). The range for these set points is 0 to 100% O2. 
42500001 REV A    123 