Instruction manual
SM425 INSTRUCTION MANUAL    9.0 CALIBRATION 
Select “Zero gas conc.” in the Calibration setup display screen. The numeric-data entry display 
then appears. Enter the desired oxygen concentration for the zero-point calibration. (The zero-
gas set ranges from 0.3 to 100 % O
2
.) 
Enter 00098 for an oxygen concentration of 0.98 vol% O
2
. 
9.2.1.4 Setting Span-gas Concentration Value 
To enter the oxygen concentration of the span-gas: 
Select “Span gas conc.” in the Calibration setup display screen. Enter the desired span-gas 
oxygen concentration using the numeric-data entry display. If instrument air is used as the span 
gas, enter 21 % O
2
. (The span-gas set ranges from 4.5 to 100 % O
2
.) 
Enter 02100 for an oxygen concentration of 21 vol% O
2
. 
Instrument air is here defined as dry air with a dew-point temperature of no higher than -20°C. If the dew-
point temperature is higher than -20°C, use a hand-held oxygen analyzer to measure the actual oxygen 
concentration. 
If using the ZO21S Standard Gas Unit and atmospheric air as the span gas, determine the actual oxygen 
concentration with a hand-held oxygen analyzer. 
- When instrument air is used for the span calibration, remove the moisture from the instrument air to a 
dew-point temperature of -20°C and remove any oil and particulate from that air. 
- Measurement accuracy is adversely affected if the air is not clean and sufficiently dehumidified. 
9.2.1.5 Setting Calibration Timing 
Calibration timing consists of five parameters: Hold time, Csl time, Interval, Start date, and Start time. 
These parameters are entered using the “Calibration Timing Display” which appears when “Timing” is 
selected on the “Calibration setup” display screen. Select the appropriate parameter and enter the 
desired value. 
Figure 9-6 “Calibration Timing” Display 
Hold time is the stabilization period following a calibration during which process gas diffuses back into the 
sensor. Normal measurements are resumed after completion of this stabilization period. See Figure 9-7. 
During this period, the analog output is held to the value obtained just prior to calibration. For more 
148 42500001 REV A 










