SM425 INSTRUCTION MANUAL    2.0 SPECIFICATIONS 
6 42500001 REV A 
±1% Maximum value of range setting; from 0 to 5 vol% O
2
 to 0 to 25 vol% O
2
 range 
 (Sample gas pressure: within ±4.9 kPa) 
±3% Maximum value of range setting; from 0 to 25 vol% O
2
 to 0 to 50 vol% O
2
 range 
 (Sample gas pressure: within ±0.49 kPa) 
±5% Maximum value of range setting; from 0 to 50 vol% O
2
 to 0 to 100 vol% O
2
 range (Sample 
gas pressure: within ±0.49 kPa) 
Drift: 
(Excluding the first two weeks in use and excluding the case where the reference air flow is 
natural convection.) 
Both zero and span ±2% Maximum value of range setting/month 
Response Time: 
Response of 90% within 5 seconds. (Measured from time when gas is introduced from calibration 
gas inlet to time when analog output starts changing in response.) 
2.2 General-use Separate-type Detector and Related 
Equipment 
General-use separate-type detector SM425P can be used in combination with the probe protector 
ZO21R-L (see Section 1.4.2). 
2.2.1 SM425P General-use Separate-type Detector 
Sample Gas Temperature: 
0 to 700°C (Probe only)  
It is necessary to mount the cell using Inconel cell-bolts when the temperature is 600°C or 
greater. 
700 to 1400°C (with High Temperature Probe Adapter)  
For high-temperature sample gas, apply 0.15 m length probe and High temperature Probe 
Adapter ZO21P-H. 
Sample Gas Pressure: 
-5 to +250 kPa (When the pressure in the furnace exceeds 3 kPa, it is recommended that you 
compensate for this pressure. When the pressure in the furnace exceeds 5 kPa, you must 
perform pressure compensation.) 
For 0.15m probe, 0.5 to 5kPa. 
No pressure fluctuation in the furnace should be allowed. 
Note: When the detector is used in conjunction with a check valve and a ZA8F Flow Setting Unit, 
the maximum pressure of sample gas is 150 kPa. When used with a check valve and a 
SM425CAL Autocalibration Unit, it is 200 kPa. If the pressure of your sample gas exceeds these 
limits, consult Teledyne Monitor Labs. 
Probe Length: 
0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m 
Probe Material: 
SUS 316 (JIS) 
Ambient Temperature: 
-20 to +150°C 