SM425 INSTRUCTION MANUAL    2.0 SPECIFICATIONS 
16 42500001 REV A 
EN 61326 Class A 
EN 55011 Class A Group 1 
EN 61000-3-2 
AS/NZS CISPR 11 
This instrument is a Class A product, and it is designed for use in an industrial environment. 
Maximum Distance between Probe and Converter: 
Conductor two-way resistance must be 10 Ω or less (when a 1.25 mm cable or equivalent is 
used, 300 m or less.) 
Construction: 
Outdoor installation, equivalent to NEMA 4X/IP66 (with conduit holes completely sealed with a 
cable gland) 
Wiring Connection: 
G 1/2, Pg13.5, M20 X 1.5 mm, 1/2 NPT, eight holes 
Installation: 
Panel, wall, or 2-inch pipe mounting 
Case: 
Aluminum alloy 
Paint Color: Door: 
Silver gray (Munsell 3.2PB7.4/1.2) 
Case: 
Silver gray (Munsell 3.2PB7.4/1.2) 
Finish: 
Polyurethane corrosion-resistance coating 
Weight: 
Approx. 6 kg 
2.4.2 Functions 
Display Functions: 
Value Display; Displays values of the measured oxygen concentration. 
Graph Display; Displays trends of measured oxygen concentration 
Data Display; Displays maintenance data, such as cell temperature, reference junction 
temperature, or maximum/minimum oxygen concentration. 
Status Message; Indicates an alarm or error condition with flashing of the corresponding icon. 
Indicates status such as warm-up or calibration by symbols. 
Alarm, Error Display; Displays alarms such as “Abnormal oxygen concentration” or errors such as 
“Abnormal cell emf.”. 
Calibration Functions: 
Auto-Calibration; Requires an Auto-calibration unit. It calibrates automatically at specified 
intervals. 
Semi-autocalibration; requires an Auto-calibration unit. Calibration is initiated either by the touch 
panel or an input control signal. 
Manual Calibration; Calibration performed by manually opening/closing the calibration gas valve. 