Instruction manual
3.0 INSTALLATION    SM425 INSTRUCTION MANUAL 
Figure 3-17 Wall Mounting 
3.6 Installation of the Calibration Gas Unit Case (E7044KF) 
The calibration gas unit case is used to store the G7001ZC zero gas cylinders. 
3.6.1 Location 
The following should be taken into consideration: 
1. Easy access for cylinder replacement 
2.  Easy access for normal operation 
3.  Close proximity to the detector and converter as well as the flow setting unit. 
4.  The temperature of the case should not exceed 40°C from exposure to direct sunlight or radiated 
heat. 
5.  Isolated from vibration 
3.6.2 Mounting 
Mount the calibration gas unit case on a pipe (nominal JIS 50A) as follows: 
1.  Use a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting the flow 
setting unit. (The sum of the calibration gas unit case and the calibration gas cylinder weighs 
approximately 4.2 kg.) 
2.  Mount the unit case on the pipe by tightening the U-bolt such that the metal fitting is firmly 
attached to the pipe. 
42500001 REV A    57 










