Table 5. Possible causes and solutions for suboptimal BLE communication
Possible Causes
Solutions
The BLE dongle orientation or location is
suboptimal.
Move or reorient the Merlin PCS slightly. Make sure
that the BLE dongle faces the implantable device.
People or objects interfere with the
communication between the Merlin PCS and
the device.
Make sure that the space between the Merlin PCS and
the device is free from interfering objects or people.
The BLE dongle is too far away from the
device.
Move the Merlin PCS closer to the device.
Other products in the vicinity are causing
electromagnetic interference (EMI).
Power off or remove equipment that could cause EMI.
Disconnecting Leads and Avoiding Shock While Handling
Disable tachyarrhythmia therapy (Enable/Disable Tachy Therapy) or program tachyarrhythmia
therapies Off during surgical implant and explant or post-mortem procedures, and when
disconnecting leads. The device can produce electrical artifacts that can be sensed by the pulse