Product Card
Table Of Contents
- Device Description
- Intended Use
- Indications
- Intended Clinical Benefits
- MRI Safety Information
- Contraindications
- Warnings
- Precautions
- Implantation Procedure
- Device Modification
- Storage and Handling
- Temperature Equilibrium
- Lead Impedance
- Suboptimal Bluetooth Communication
- Disconnecting Leads and Avoiding Shock While Handling
- External Equipment for Arrhythmia Induction
- Antiarrhythmic Drugs
- Sterilization
- Damaged Package
- Environmental Hazards
- Hospital and Medical Environments
- Environmental and Medical Therapy Hazards
- Home and Industrial Environments
- Transcutaneous Electrical Nerve Stimulation
- Radiofrequency Ablation
- Electronic Article Surveillance (EAS)
- Metal Detectors
- Cellular Phones
- Adverse Events
- Pulse Generator Header
- Sensing
- Radiopaque Identification
- Implanting the Pulse Generator
- Managing and Following Patients
- Device Longevity
- Using a Magnet
- Explanting the Pulse Generator
- Out-of-Service/Explant/Patient Death Form
- Technical Support
- Additional Information
- Physical Specifications
- Battery Information
- Device Configurations
- Inductive Communication
- BLE Operating Frequencies
- Spare Parts and Accessories
- Detection Performance in the Presence of Electromagnetic Interference in Differential Mode
- Symbols
47
Table 12. High-voltage waveforms
1
for 36 J devices
Energy Setting
Max
Min
Mean
Biphasic
Delivered pulse energy (J) (first shock)
30.7
0.1
14.8
First phase 27.2 0.1 13.2
Second phase 3.5 0.0 1.6
Delivered pulse energy (J) (sequential
shock)
35.9 0.1 17.5
First phase 31.7 0.1 15.6
Second phase 4.2 0.0 1.9
Peak ICD output voltage (V) (first
shock)
771 47 540
First phase 771 47 540
Second phase 263 13 184