Product Card
Table Of Contents
- Device Description
- Intended Use
- Indications
- Intended Clinical Benefits
- MRI Safety Information
- Contraindications
- Warnings
- Precautions
- Implantation Procedure
- Device Modification
- Storage and Handling
- Temperature Equilibrium
- Lead Impedance
- Suboptimal Bluetooth Communication
- Disconnecting Leads and Avoiding Shock While Handling
- External Equipment for Arrhythmia Induction
- Antiarrhythmic Drugs
- Sterilization
- Damaged Package
- Environmental Hazards
- Hospital and Medical Environments
- Environmental and Medical Therapy Hazards
- Home and Industrial Environments
- Transcutaneous Electrical Nerve Stimulation
- Radiofrequency Ablation
- Electronic Article Surveillance (EAS)
- Metal Detectors
- Cellular Phones
- Adverse Events
- Pulse Generator Header
- Sensing
- Radiopaque Identification
- Implanting the Pulse Generator
- Managing and Following Patients
- Device Longevity
- Using a Magnet
- Explanting the Pulse Generator
- Out-of-Service/Explant/Patient Death Form
- Technical Support
- Additional Information
- Physical Specifications
- Battery Information
- Device Configurations
- Inductive Communication
- BLE Operating Frequencies
- Spare Parts and Accessories
- Detection Performance in the Presence of Electromagnetic Interference in Differential Mode
- Symbols
49
Table 13. High-voltage waveforms
2
for 40 J devices
Energy Setting
Max
Min
Mean
Delivered pulse energy (J) (sequential
shock)
35.6 0.1 15.0
Peak ICD output voltage (V) (first shock) 850 45 580
Peak ICD output voltage (V) (sequential
shock)
890 45 580
Biphasic
Delivered pulse energy (J) (first shock) 36.0 0.1 17.5
First phase 32.1 0.1 15.6
Second phase 3.9 0.0 1.9
Delivered pulse energy (J) (sequential
shock)
40.0 0.1 17.5
First phase 35.6 0.1 15.6
Second phase 4.4 0.0 1.9