Instruction manual

3-19
Operation
Action Result
1. Turn the key to the
ZONAL position.
Open-door operation up to 3000 rpm is enabled. JCF-Z appears in the
Rotor: field.
2. Install the rotor. Leave the
door open when installation
is complete.
3. Press [
ROTOR]. Press the
JCF-Z soft key. Enter the
remaining run parameters.
4. Press [
ENTER], then
[
START].
The rotor begins accelerating to 2000 rpm, which is the default
loading speed. To select a different loading speed, enter a speed
of 3000 rpm or less before or after the rotor starts to spin.
If an
ω
2
t run is specified,
ω
2
t starts accumulating.
5. Load the rotor.
6. Close the door. The rotor accelerates to the set speed. If you entered a loading speed
other than 2000 rpm, re-enter the run speed after you close the door.
7. End the run. When the run is complete (the run time has expired, the set
ω
2
t value
has been reached, or [
STOP] has been pressed), the rotor decelerates to
the default unloading speed of 2000 rpm. If desired, you may enter a
different unloading speed of 3000 rpm or below after the rotor
reaches 2000 rpm.
ZONAL
Accel : MAX Decel : MAX
Rotor : JCF-Z
Speed : 0 RPM = x G
Time : 00:00
Temp : 25°C Max Temp : 29°C
JCF-Z
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Accel : MAX Decel : MAX
Rotor : JCF-Z
Speed : 20 000 RPM = x G
Time : 00:45
Temp : 20°C Max Temp : 24°C
SPEED
TIME
TEMP
A/D
OPT
ROTOR
JA
JS
JLA
JV
JE
JCF-Z