Instruction For Use Avanti J-26XP High-Performance Centrifuge J326XP-IM-5AB January 2012 Beckman Coulter, Inc. 250 S. Kraemer Blvd.
Avanti J-26XP High-Performance Centrifuge J326XP-IM-5AB (January 2012) Copyright © 2012 Beckman Coulter, Inc. All rights reserved. No part of this document may be reproduced or transmitted in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission from Beckman Coulter, Inc. Find us on the World Wide Web at: www.beckmancoulter.
Safety Notice Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate the instrument. Do not attempt to perform any procedure before carefully reading all instructions. Always follow product labeling and manufacturer’s recommendations. If in doubt as to how to proceed in any situation, contact your Beckman Coulter Representative.
Safety Notice Electrical Safety Electrical Safety To reduce the risk of electrical shock, this instrument uses a three or five-wire electrical cord and plug to connect this equipment to earth-ground. To preserve this safety feature: • Make sure that the matching wall outlet receptacle is properly wired and earth-grounded. Check that the line voltage agrees with the voltage listed on the name-rating plate affixed to the centrifuge. • Never use a three-to-two wire plug adapter.
Safety Notice Chemical and Biological Safety procedures and methods to prevent spread of disease. Because spills may generate aerosols, observe proper safety precautions for aerosol containment. Do not run toxic, pathogenic, or radioactive materials in this centrifuge without taking appropriate safety precautions.
Safety Notice Chemical and Biological Safety vi J326XP-IM-5AB
Contents Safety Notice, iii Alerts for Danger, Warning, Caution, and Note, iii Safety During Installation and/or Maintenance, iii Electrical Safety, iv Safety Against Risk of Fire, iv Mechanical Safety, iv Chemical and Biological Safety, iv Introduction, xiii Certification, xiii Scope of This Manual, xiii Conventions, xiv CFC-Free Centrifugation, xiv Radio Interference, xiv Recycling Label, xv CHAPTER 1: Specifications, 1-1 Control Features, 1-1 Operational Features, 1-2 Physical Data, 1-2 Available Rot
Contents CHAPTER 2: Centrifuge Function and Safety Features, 2-1 Centrifuge Function, 2-1 Safety Features, 2-2 Housing and Door, 2-2 Rotor Chamber, 2-2 Drive, 2-2 Friction Reduction System (FRS), 2-3 Temperature Sensing and Control, 2-3 Overtemp System, 2-3 Name Rating Plate, 2-3 Controls and Indicators, 2-3 Power Switch, 2-3 Control Panel, 2-4 CHAPTER 3: Introduction, 3-1 Summary of Avanti J-26XP Run Procedures, 3-2 Section and Loading, 3-3 Entering Run Parameters, 3-4 Selecting a Rotor, 3-4 Setting Ru
Contents Storage and Transport, 5-5 Storage, 5-5 Returning a Centrifuge, 5-5 Supply List, 5-6 Replacement Parts and Supplies, 5-6 APPENDIX A: Introduction, A-1 Avanti® J Series Centrifuge Warranty, Warranty-1 ix
Illustrations x 1.1 Single-Phase Electrical Requirements, 1-9 1.2 Three-Phase “Y” Electrical Requirements, 1-9 1.3 Correct and Incorrect Three-Phase Power Configurations, 1-10 1.4 Rear View and Dimensions, 1-12 1.5 Checking the Rotor for Drive Pins, 1-13 2.1 The Power Switch, 2-4 2.2 The Control Panel, 2-4 3.1 Acceleration and Deceleration Settings Graph, 3-8 3.2 Elutriation Controls, 3-9 4.1 Emergency Door Release Latch Access, 4-5 4.2 Manual Door Release, 4-6 4.
Tables Tables 1.1 Nominal Supply Voltage Ratings for the Avanti J-26XP, 1-8 1.2 Required Wire Connections, 1-10 3.1 Acceleration Settings, 3-7 3.2 Deceleration Settings, 3-8 4.
Tables xii
Introduction Certification To ensure full system quality, Beckman Coulter Avanti® J-26XP centrifuges are manufactured in a registered ISO 9001 or 13485 facility. They have been designed and tested to be compliant (when used with Beckman Coulter rotors) with the laboratory equipment requirements of applicable regulatory agencies. Declarations of conformity and certificates of compliance are available at www.beckmancoulter.com.
Introduction Conventions Conventions Certain symbols are used in this manual to call out safety related and other important information. These international symbols may also be displayed on the centrifuge and are reproduced on the inside back cover of the manual. Typographic Conventions Certain typographic conventions are used throughout this manual to distinguish names of user interface components, such as keys and displays. • Control knob names (for example, ROTOR and SPEED) appear in bold type.
Introduction Recycling Label Recycling Label A28219-AA This symbol is required in accordance with the Waste Electrical and Electronic Equipment (WEEE) Directive of the European Union. The presence of this marking on the product indicates: 1. The device was put on the European market after August 13, 2005 and 2. The device is not to be disposed via the municipal waste collection system of any member state of the European Union.
Introduction Recycling Label xvi J326XP-IM-5AB
CHAPTER 1 Specifications and Preinstallation Requirements Specifications Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data, without guarantee.
Specifications and Preinstallation Requirements Specifications Specifications Description Acceleration maximum or slow Deceleration maximum, slow, or off a. During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. Refer to applicable rotor manuals for specific rotor operating range information. b. To reach temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber during operation.
Specifications and Preinstallation Requirements Available Rotors Specification Description Electrical Requirements • 200/208/240-V, single-phase instrument — 180–264 VAC, 30 A, 50/60 Hz • 230-V, single-phase instrument — 180–264 VAC, 30 A, 50 Hz • 220/380-V plus neutral, three-phasea instrument — 313–457 VAC plus neutral, 16 A, 50 Hz Electrical Supply Class I Maximum Heat dissipation into Room Under Steady-State Conditions 2.
Specifications and Preinstallation Requirements Available Rotors Rotor Profile 1-4 Rotor Code Max RPMa JA-25.15 Ti Fixed Angle, 25° (24 place)d rmax = 106 mm (outer row) rmax = 86 mm (inner row) 25.15 25 000 JA-21 Fixed Angle, 40° (18 place) rmax = 102 mm 21 21 000 JA-20.1 Fixed Angle, 23° (32 Place) rmax = 115 mm (outer row) rmax = 98 mm (inner row) 20.1 20 000 JA-20 Fixed Angle, 34° (8 place) rmax = 108 mm 20 JA-18.
Specifications and Preinstallation Requirements Available Rotors Rotor Profile J326XP-IM-5AB Rotor Code Max RPMa Max RCF ( g) Max Capacity Rotor Manual Number JA-14 Fixed Angle, 25° (6 place) rmax = 137 mm 14 14 000 30 100 6 250 mL J-TB-004 F14BCI-14x50cy Fixed Angle, 34° (14 place) rmax = 153 mm F50C 14 000 33 500 14 50 mL — F14BCI-6x250y Fixed Angle, 23° (6 place) rmax = 134 mm F250 14 000 30 000 6 250 mL — JA-12 Fixed Angle, 35° (12 place) rmax = 144 mm 12 12 000 23 20
Specifications and Preinstallation Requirements Available Rotors Rotor Profile 1-6 Rotor Code Max RPMa Max RCF ( g) Max Capacity Rotor Manual Number JS-13.1 Swinging Bucket (6 place) rmax = 140 mm 13.1 13 000 26 500 6 50 mL J-TB-036 JS-24.38 Swinging Bucket (6 place) rmax = 161 mm 24.38 10 000e 18 000 6 38.5 mL J-TB-058 JS-24.15 Swinging Bucket (6 place) rmax = 171.3 mm 24.15 10 000e 19 200 6 15 mL J-TB-058 JS-7.5 S Swinging Bucket (4 place) rmax = 165 mm 7.
Specifications and Preinstallation Requirements Preinstallation Requirements Rotor Profile Rotor Code Max RPMa Max RCF ( g) JCF-Z Continuous Flow/ Zonal Rotorf JCFZ 20 000 39 900 JE-5.0 Elutriation Rotor rmax large chamber = 168 mm standard chamber = 125 mm Sanderson chamber = 126 mm 5.0 5 000 Description Max Capacity 660 mL (standard core) 1250 mL (large core) 240 mL (small core) Rotor Manual Number JCFZ-IM JE5-IM 4 700 40 mL 3 500 4.0 mL 4 230 5.5 mL a.
Specifications and Preinstallation Requirements Preinstallation Requirements • Power receptacle (see Figure 1.1 or Figure 1.2) • Drill for drilling holes in the floor for installation of anchoring kit bolts (see Securing the Centrifuge to the Floor, later in this section). A 9.5 mm (3/8-inch) drill is required for concrete floors. A 6.4 mm (1/4-inch) drill is required for wood floors.
Specifications and Preinstallation Requirements Preinstallation Requirements Single and Three-Phase Power Connections Figure 1.1 shows the power connection for single-phase centrifuges, including earth-ground and two power leads with 30-ampere circuit breakers. Figure 1.2 shows the power connection for threephase centrifuges, including earth-ground and three power leads with minimum 16-ampere circuit breakers and common neutral connection. Table 1.2 contains wiring information. Figure 1.
Specifications and Preinstallation Requirements Preinstallation Requirements Table 1.2 Required Wire Connections Symbol Wire Insulation Color Terminal Harmonized North American Green/Yellow Earth ground Light Blue Neutral N L Brown Live or Line L L Black (three-phase connections only) Live or Line (qty 2) L — Additional Requirements for Three-Phase Power Connections For three-phase power service to the centrifuge, use the “Y” connected configuration shown in Figure 1.3.
Specifications and Preinstallation Requirements Preinstallation Requirements Space and Location Requirements 1 Locate the instrument on a clean, level floor. 2 Select a location away from heat-producing laboratory equipment. • If ambient temperature exceeds 38°C (100°F), premature component failure may result. 3 In addition to space for the centrifuge, allow a 7.7-cm (3-in.) clearance on each side of the centrifuge and a 16-cm (6.25-in.) clearance behind the centrifuge for air circulation.
Specifications and Preinstallation Requirements Preinstallation Requirements Figure 1.4 Rear View and Dimensions cm in. 86 34 149 58.5 1 116 45.5 2 1.8 m 6.0 ft. 86 34 3 4 16.0 6.25 5 71 28 1. Air Exit 2. Power Cable 3. Air Diverter 4. Depth 5. Air Intake Bio-Safety Level 3 Installation For laboratories with epoxy aggregate (resinous) floors, such as BSL-3 labs, a non-invasive installation kit (PN 393316) is available.
Specifications and Preinstallation Requirements Using J2 Series Rotors in the Avanti J Using J2 Series Rotors in the Avanti J CAUTION Do not use the Beckman Coulter JA-10, JS-7.5, JA-14, or JS-13 rotors in the Avanti J-26XP centrifuge before reading this information. Checking for Rotor Drive Pins Rotors used in Avanti J series centrifuges must have drive pins in the rotor drive hole (see Figure 1.5).
Specifications and Preinstallation Requirements Using J2 Series Rotors in the Avanti J The rotor pins are positioned parallel to the Beckman name engraved at the center of the rotor body (see Figure 1.5). Knowing the pin orientation before you install the rotor will help to ensure that you position the rotor properly on the hub, minimizing the chance of hub damage.
Specifications and Preinstallation Requirements Using J2 Series Rotors in the Avanti J The following information pertains to JCF-Z rotors manufactured before March, 1997, and to JCF-Z rotors that have not been modified at the factory. • When the bearings in the JCF-Z rotating seal assembly get old or worn, the resulting “drag” on the bearings changes the rotor’s dynamic properties, making them similar to those of several Beckman Coulter fixed angle rotors.
Specifications and Preinstallation Requirements Using J2 Series Rotors in the Avanti J 1-16 J326XP-IM-5AB
CHAPTER 2 Description This section describes the Avanti J-26XP centrifuge components and their functions. It also describes system safety features and centrifuge controls and displays. Refer to the applicable rotor manual for rotor descriptions. Centrifuge Function and Safety Features Centrifuge Function The Avanti J-26XP is a refrigerated centrifuge that generates centrifugal forces required for a wide variety of applications.
Description Centrifuge Function and Safety Features The user interface has a digital display, which can be toggled to show either actual centrifuge values or user-entered setup values, knobs for parameter entry, and touch keys. Diagnostic messages on the control panel will light to alert the operator to conditions that may need attention. Safety Features Avanti J-26XP centrifuges have been designed and tested to operate safety indoors at altitudes up to 2 000 m (6 562 ft).
Description Name Rating Plate Friction Reduction System (FRS) The friction reduction system (FRS) uses a mechanical rotary vane vacuum pump to reduce chamber pressure to approximately one-quarter atmosphere (190 mm Hg). The pump turns on after the run is started, before rotor friction reaches a high level. When the required vacuum level is reached, the pump turns off. Vacuum in the chamber is vented during rotor deceleration.
Description Controls and Indicators Figure 2.1 The Power Switch Control Panel The control panel (Figure 2.2) is mounted at an angle on the centrifuge top rear for easy visibility and access. It contains a digital display for actual and setup values, knobs and touch keys for parameter entry, and additional touch keys for system control. Figure 2.2 The Control Panel 1 1. Digital Display Digital Display The display shows current status of the instrument whenever the power is turned on.
Description Controls and Indicators • When HOLD is selected (no run time is specified), the TIME display shows the time elapsed since START was pressed. After 3 hours (180 minutes) have elapsed, the timer will reset to 0 and continue counting elapsed time until STOP is pressed. Knobs and Touch Keys The rotor name, speed, time, and temperature settings are entered by turning the appropriate knob until the required value is displayed.
Description Controls and Indicators TIME Selection Two time modes are available. The HR:MIN mode is used for runs of specified lengths. The HOLD mode is used for continuous runs of unspecified lengths. The HOLD key is used to toggle between the two modes. In HR:MIN mode: • The TIME knob is used to set run time, which can be set up to 3 hours and 0 minutes (180 minutes). • Each TIME knob increment is 1 minute. A full turn of the knob equals 60 minutes. • The time setting can be changed during a run.
Description Controls and Indicators System Keys START Pressed to begin a run. When START is pressed, the display immediately shows the actual centrifuge values. The green START light flashes during acceleration until set speed is reached, and then remains on continuously until deceleration begins. STOP Pressed to end a run; the centrifuge decelerates to a complete stop according to the deceleration setting. The red STOP light blinks during deceleration.
Description Controls and Indicators 2-8 J326XP-IM-5AB
CHAPTER 3 Operation Introduction This section contains detailed centrifuge operating procedures. A summary is provided on page 3-2. If you are an experienced user of this centrifuge, you can turn to the summary for a quick review of operating steps. WARNING Normal operation may involve the use of solutions and test samples that are pathogenic, toxic, or radioactive. Handle body fluids with care because they can transmit disease.
Operation Summary of Avanti J-26XP Run Procedures Summary of Avanti J-26XP Run Procedures For runs at temperatures other than ambient, precool or prewarm the rotor to the required temperature before the run. 1 Press the POWER switch to on (I). Open the centrifuge door. 2 Install the rotor. Make sure that the rotor load is balanced and that the rotor is securely fastened to the centrifuge drive hub. Close the centrifuge door.
Operation Section and Loading Section and Loading For fast temperature equilibration, precool or prewarm the rotor to the required temperature before the run. 1 Turn the power switch on (I). • Power is applied to the system, and the display illuminates. 2 Step on the foot pedal to open the door. • The door opens. 3 Install the rotor according to the instructions in the rotor manual. a. Always run the rotor with a balanced load. b.
Operation Entering Run Parameters Entering Run Parameters Selecting a Rotor 1 Turn the ROTOR knob until the rotor in use appears in the display. • Small green LEDs by each rotor type abbreviation light one at a time, and the rotors of that type appear consecutively in the display. ROTOR JA JS JLA JSP JV JE JCF Setting Run Speed Enter a run speed up to 26 000 rpm, or up to the maximum speed of the rotor in use.
Operation Entering Run Parameters 3 Turn the SPEED knob until the required RPM appears in the SPEED display. • When the run is started, the centrifuge operates at the set speed. • The corresponding RCF is automatically calculated. a. Press RPM/RCF to display the RCF value. Setting RCF 1 Check to see if the red RCF light is lit on the display. a. If it is lit, go to step 3. 2 Press the RPM/RCF key to select the RCF mode. • On the display, the red RCF ( g) LED lights.
Operation Entering Run Parameters Timed Run 1 Turn the TIME knob until the required run time appears in the TIME display. • If you enter more than 59 minutes in the minutes field, the system automatically converts the entry to hours and minutes after ENTER is pressed. 2 When all run parameters are entered, press START. • The rotor begins to accelerate and the display begins showing the remaining run time. • The run will end when the time counts down to 0 or when STOP is pressed.
Operation Entering Run Parameters Setting Run Temperature Select a run temperature between –10 and +40°C, in 1°C increments. NOTE For runs at temperatures other than ambient, always refrigerate or warm the rotor to the required run temperature before the run. For low-temperature runs, precool the system by running a precooled rotor at 2000 rpm at the required temperature for at least 30 minutes. 1 Turn the TEMP°C knob until the required temperature appears in the display.
Operation Entering Run Parameters Table 3.2 Deceleration Settings Maximum Deceleration Slow Deceleration Off Full brake is used from set speed to near 0 rpm, to bring the rotor to a stop in the shortest possible time. Reduced brake is used during the last few rpm until 0 is reached, to minimize disturbance of the sample. Reduced torque is used from set speed to 500 rpm, causing deceleration to 500 rpm to take twice as long as maximum deceleration.
Operation Entering Run Parameters 1 Press the ACCEL key to toggle between the MAX and SLOW acceleration settings. • The selected setting lights. MAX SLOW ACCEL Setting the Deceleration Rate 1 Press the DECEL key to toggle between the MAX, SLOW, and OFF deceleration settings. • The selected setting lights. MAX SLOW OFF DECEL Setting Up an Elutriation Run Centrifuges equipped for elutriation have elutriation controls as shown in Figure 3.
Operation Entering Run Parameters 1 Install the rotor. 2 Set up sample and buffer reservoirs and tubing lines as described in the rotor manual. a. Route the tubing lines through the elutriator port holes at the left side of the door seal. b. Close the centrifuge door. 3 Turn the ROTOR knob to the JE position. • 5.0 appears in the ROTOR ID display. ROTOR JA JS JLA JSP JV JE JCF 4 Enter the remaining run parameters. Refer to the JE-5.0 rotor manual for rotor setup and operating instructions.
Operation Entering Run Parameters 7 When the rotor speed has stabilized, look through the port in the centrifuge door and turn the DELAY knob until the elutriation chamber in the rotor is synchronized with the strobe. • (The chamber will appear to be motionless.) • Sample can now be injected into the lines. • See the rotor instruction manual for detailed instructions. DELAY ELUTRIATOR 1 1.
Operation Changing Parameters During a Run Changing Parameters During a Run While a run is in progress, the speed, time, temperature, and acceleration or deceleration setting can be changed without stopping the run. The rotor name parameter cannot be changed during a run: Changing Time Settings 1 To increase or decrease the remaining run time, turn the knob to the new run time.
Operation Unloading the Centrifuge 1 Press STOP to initiate deceleration. • The light on the STOP key flashes until the rotor comes to a stop. 2 After the rotor stops spinning, step on the foot pedal to open the door. NOTE During runs at very cold temperatures, nominally –10°C or below, ice may form around the door opening, causing the door to stick shut. If this happens, press firmly on the door in several locations around the door opening and depress the foot pedal again.
Operation Unloading the Centrifuge 3-14 J326XP-IM-5AB
CHAPTER 4 Troubleshooting Introduction This chapter lists possible malfunctions, along with probable causes and corrective actions. Maintenance procedures are given in CHAPTER 5, Care and Maintenance. For any problems not covered here, contact Beckman Coulter Field Service (1-800-742-2345 in the United States for assistance. NOTE It is your responsibility to decontaminate the instrument, as well as any rotors and/or accessories, before requesting service by Beckman Coulter Field Service.
Troubleshooting User Messages Table 4.1 contains information on causes and recommended actions for each diagnostic condition. If you are unable to correct the problem, call Beckman Coulter Field Service (in the United States, call 1-800-742-2345; outside of the U.S., call your local Beckman Coulter office).
Troubleshooting User Messages Table 4.1 Diagnostic Message Chart (Continued) Diagnostic Message Problem Result Recommendation System-identified rotor does not match userset rotor; set speed above rotor maximum System lowers run speed to the maximum of the rotor identified by system Press CLEAR to stop LED from flashing. System cannot identify rotor (rotor ID system malfunction) Run shuts down with maximum brake Check rotor and restart run.
Troubleshooting User Messages Table 4.1 Diagnostic Message Chart (Continued) Diagnostic Message Result Recommendation Momentary power failure; rotor does not come to a complete stop Run continues when power resumes; elapsed time continues counting from current value Press CLEAR to stop LED from flashing. Power failure; rotor comes to a complete stop Run restarts when power resumes; elapsed time starts counting from 0 Press CLEAR to stop LED from flashing.
Troubleshooting Accessing the Rotor in Case of Power Failure Accessing the Rotor in Case of Power Failure WARNING Any maintenance procedure requiring removal of a panel exposes the operator to the possibility of electrical shock and/or mechanical injury. therefore, turn the power off (O) and disconnect the instrument from the main power source, and refer such maintenance to qualified service personnel.
Troubleshooting Accessing the Rotor in Case of Power Failure 1 Turn the power switch to off (O) and unplug the power cord from the main source. 2 Insert a 4-mm or 5/32-in. Allen wrench straight through one of the holes (either one) and turn the wrench to the left (counterclockwise) about one-half turn. • The latch disengages. 3 Repeat step 2, inserting the Allen wrench through the other hole. • After the second latch disengages, the front panel will fall forward about an inch from the top.
Troubleshooting Accessing the Rotor in Case of Power Failure 5 Pull the interlock lever out and to the left (at about a 45-degree angle), and while holding it out, step on the foot pedal. • Depending on the level of vacuum in the chamber at the time of the power failure, the door may or may not open. a. If it opens, go to step 6. b. If it does not open, you will need to manually vent the vacuum. Go to step 7. 6 After the door opens, first release the foot pedal and then release the interlock lever.
Troubleshooting Accessing the Rotor in Case of Power Failure Figure 4.3 Venting the Chamber Vacuum 5 4 6 7 1. 2. 3. 4. Step 1 Step 2 Step 3 Vacuum Pump 5. Grasp red hose and pull up Do not disconnect the tubing lines tied to the red hose. 6. Red hose removed from fitting 7.
Troubleshooting JCF-Z Rotor Identification 2 Hold the left top edge in place and insert the Allen wrench through the hole. a. Turn to the right (clockwise) about one-half turn, until you reach a stop. • The latch engages. 3 Repeat step 2 on the right side • The second latch engages. Figure 4.4 Front Panel Retaining Strip 1 1. Retaining Strip JCF-Z Rotor Identification The rotor identification system can, under two specific conditions, misidentify the JCF-Z continuous flow/zonal rotor.
Troubleshooting JCF-Z Rotor Identification 1 On the first few uses of a new JCF-Z rotor, or in an older JCF-Z rotor with newly replaced bearings, run the rotor from 0 to 5000 rpm and back to 0 rpm three times. • This procedure will distribute lubricant around the bearings in the rotating seal assembly, reducing drag.
CHAPTER 5 Care and Maintenance Introduction This section describes routine care and maintenance procedures that you should perform regularly or as required. For maintenance not covered in this manual, contact Beckman Field Service.* Refer to the applicable rotor manual and Rotors and Tubes for J Series Centrifuges (publication JR-IM) for instructions on the care of rotors and accessories.
Care and Maintenance Maintenance 2 Check the air filter on the back panel for obstructions. a. Keep vents clear and clean. 3 Wipe condensation out of the chamber between runs with a sponge or clean cloth to prevent chamber icing. 4 If chamber icing occurs, defrost the system and wipe moisture out of the chamber before use. a. To defrost the system, set the temperature to 30°C for 20 minutes. • (These are suggested settings that may be adjusted as appropriate for your laboratory conditions.
Care and Maintenance Maintenance WARNING Be careful when examining or cleaning the chamber and chamber gasket, as sharp glass fragments may be embedded in their surfaces. 1 Examine the chamber gasket to make sure that no glass particles are retained in it. a. 2 Carefully remove any glass particles that may remain. Carefully wipe away any glass particles that remain in the chamber.
Care and Maintenance Maintenance Replacing the Air Filter 1 Check the air filter regularly and replace it about once a year, or more often if it looks dirty. • The air filter is not fastened to the centrifuge, so no tools are required for removal or installation. 1 4 2 3 1. Frame Edge 2. Air Filter 2 3. Retaining Strip 4. Frame Edge To remove the air filter, hold the side edges and lift the filter straight up until the bottom edge is above the centrifuge bottom retaining strip. a.
Care and Maintenance Circuit Breaker and Fuses Circuit Breaker and Fuses There are no user-replaceable fuses in the Avanti J-26XP centrifuge. If the centrifuge circuit breaker trips for any reason, the power switch will move to the OFF (O) position. Reset the circuit breaker by turning the power switch back to the ON (I) position. If it trips again immediately, do not reset it. Call Beckman Coulter Field Service.
Care and Maintenance Supply List Supply List Refer to the High Performance, High Speed, High Capacity Rotors, Tubes & Accessories catalog (BR-8102, available at www.beckmancoulter.com) and the applicable rotor manual for materials and supplies needed for rotors. For your convenience, a partial list of centrifuge supplies is given below. Replacement Parts and Supplies Description Air filter 885218 Beckman Solution 555™ (1 qt) 339555 Zonal Bracket Kit for Cole-Parmer size 16 tubing (6.4-mm [1/4-in.] O.
APPENDIX A Temperature Calibration Procedure Introduction The Avanti J-26XP centrifuge specification for temperature control is 2°C of the set temperature. That means that your sample will stay within 2°C of set temperature at all times, after rotor and system equilibration, described below. (During transient conditions, such as acceleration and deceleration, the rotor temperature may be outside this range.
Temperature Calibration Procedure Introduction 6 If the measured and set temperatures are different, note how many degrees different they are and adjust the set temperature up or down that number of degrees.
Beckman Coulter, Inc. Avanti® J Series Centrifuge Warranty Subject to the exceptions and upon the conditions specified below Beckman Coulter, Inc.
Beckman Coulter, Inc.
Symbol Symbol Symbole Símbolo Simbolo Title / Titel / Titre / Titulo / Titolo Dangerous voltage Gefährliche elektrische Spannung Courant haute tension Voltaje peligroso Pericolo: alta tensione ! Attention, consult accompanying documents Achtung! Begleitpapiere beachten! Attention, consulter les documents joints Atención, consulte los documentos adjuntos Attenzione: consultare le informazioni allegate On (power) Ein (Netzverbindung) Marche (mise sous tension) Encendido Acceso (sotto tensione) O
www.beckmancoulter.com © 2012 Beckman Coulter, Inc.