Specifications
PN SJ-IM-8AC
1-7
Description
Controls and Indicators
1
The keypad is used to enter numerical run parameters, to enter or recall a program
number, and to select acceleration and/or deceleration rates. In addition to keys 0
through 9, the keypad includes plus or minus/decimal key and a CE (clear entry) key.
The ± key is used to enter a temperature setting below 0°C (for example, –2°C), or to
enter a decimal point in a rotor entry code, depending on the parameter key selected.
The CE (clear entry) key clears any parameter on the display. Pressing CE will also clear
some diagnostic messages. If a diagnostic message does not clear when CE is pressed,
see CHAPTER 3, Troubleshooting.
The ENTER key can be pressed to save parameter inputs to system memory during
manual run setup, and to start the centrifuge (ENTER then START must be pressed to
start the centrifuge).
Pressed during a run to display set values for 5 seconds.
Pressing ENTER, then START begins the run. (START must be pressed within
5 seconds of pressing ENTER.) This key combination can also be used to abort a
deceleration process and restart the centrifuge.
Can be pressed at any time while the rotor is spinning to terminate a run. The
centrifuge decelerates to a complete stop according to the deceleration setting. The
green light next to the key flashes while the rotor decelerates. An audio indicator
sounds when the rotor comes to a complete stop.
Locks and unlocks the door. (The instrument will not accept the open command if the
rotor is spinning.)