Specifications

PN SJ-IM-8AC
2-3
Operation
Preparation
2
6
Press TIME, then use the keypad to enter the run time (to 99 hrs, 59 minutes); or press TIME
twice for a hold (continuous) run.
7
Press TEMP°C, then use the keypad to enter the required run temperature (10 to +40°C).
8
Press ACCEL, then use the keypad to enter the selected acceleration rate number, 1 or 2.
9
Press DECEL then use the keypad to enter the selected deceleration rate number, 1, 2, or 3.
10
Check that all parameters are correct and that the door is closed. Press ENTER then press START
(within 5 seconds).
11
Wait for the run to end, or end the run by pressing STOP.
12
When the rotor stops (a tone sounds), press DOOR to unlock the chamber door; lift the door
open to remove the rotor.
Preparation
Prepare the rotor for centrifugation as described in the applicable rotor manual. For runs at other
than room temperature, refrigerate or warm the rotor beforehand for fast equilibration.
Installing the Rotor
The power must be turned on before you can unlock and open the chamber door. (To end a run for
any reason, do not turn the power switch off; press
STOP instead.)
1
Turn the power switch on (I).
Indicator lights on the control panel light up.
2
Press DOOR to unlock the door.
The instrument will accept this command only when the rotor is at rest.
3
Lift the door up to open.