Specifications

PN SJ-IM-8AC
3-5
Troubleshooting
Accessing the Rotor in Case of Power Failure
3
2
Insert the hex wrench (729317) or a similar tool into the openings on each side of the centrifuge
(see Figure 3.1) to release the top front panel.
CAUTION
LISTEN CAREFULY. Do not proceed if any sound or vibration is coming from
the drive.
Figure 3.1 Manual Door Release
3
Lower the panel as far as the tether cords allow.
CAUTION
Do not attempt to remove tether cords. They are required to assure safe operation
of the centrifuge.
4
Use the latch override key (368247) to turn the right-hand latch bolt to the right (clockwise) and
the left-hand latch bolt to the left (counterclockwise) until the latches release.
If the rotor is still spinning, close the door and turn the latches to secure it.
Wait until the rotor stops spinning to repeat this step.
1. Hex Wrench 2. Latch Override Key
TM
1
1
2