Specifications

PN SJ-IM-8AC
4-2
Care and Maintenance
Maintenance
2
Check the air filter on the back panel for obstructions.
Keep vents clear and clean.
3
Wipe condensation out of the chamber between runs with a sponge or clean cloth to prevent
chamber icing.
4
If chamber icing occurs, defrost the system and wipe moisture out of the chamber before use.
To defrost the system, set the temperature to 30°C for 20 minutes.
(These are suggested settings that may be adjusted as appropriate for your laboratory
conditions.)
NOTE
Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed method will not damage the
equipment.
Cleaning
Clean the centrifuge frequently. Always clean up spills when they occur to prevent corrosives or
contaminants from drying on component surfaces.
1
To prevent accumulations of sample, dust, and/or glass particles from broken sample tubes,
keep the chamber clean and dry by frequent wiping with a cloth or paper towel.
For thorough cleaning, wash the chamber using a mild detergent such as Solution 555.
Dilute the detergent with water (10 parts water to 1 part detergent).
Rinse thoroughly and dry completely.
2
Clean the centrifuge exterior surfaces by wiping with a cloth dampened with Solution 555.
Dilute the detergent with water (10 parts water to 1 part detergent).
Do not use acetone.
3
Clean the drive hub regularly using Solution 555 and a soft brush.
Dilute the detergent with water (10 parts water to 1 part detergent).
Rinse thoroughly and dry completely.