Technical information

Centrifuge Standard Operating Procedure 19
8. PREVENTATIVE MAINTENANCE
Users are not to perform maintenance. These procedures are carried out by
the Instrumentation Technician.
8.1 Daily
- Check the centrifuge interior for condensation
- Check the log book for any problems or concerns
8.2 Weekly
- Clean interior of centrifuge and drive hub with mild detergent or diluted
Beckman 555 solution
- Check rotors for discoloration
- Grease the o-rings, gaskets and threads of the centrifuge, rotors, and
canisters
8.3 Monthly
- Clean and lubricate the rotor pins and bucket pin sockets
8.4 Three Months
- Check the vacuum pump oil level and clarity
8.5 Six Months
- Replace rotor o-rings and centrifuge gaskets if necessary
8.6 Annually
- Check and install new air filter if required