Arm Type Bluetooth Blood Pressure Monitor Model No.: BPM63ZB INSTRUCTION MANUAL PLEASE READ THIS INSTRUCTION MANUAL COMPLETELY BEFORE OPERATING THIS UNIT.
Index Intended Use 1 Important Information Before Use 2 Product Identification 3 Description of LCD Display 3 Battery Installation 4 Setting the Date and Time 5 Placement of the Pressure Sleeve 6 Measurement of Pulse Rate and Blood Pressure 8 Irregular Heartbeat Detector 10 World Health Organization (WHO) 10 Memory Function 11 Operating the Bluetooth function 12 Troubleshooting 13 Error Codes 13 Care and Maintenance 14 Applied Standards 14 Technical Specification 14 FCC St
Intended Use BPM63ZB Series Blood Pressure Monitor applies the Oscillometric method to measure human systolic / diastolic blood pressure and heart rate.All values are shown on the LCD monitor. This device is designed only for adults.
Important Information Before Use 1. Blood pressure measurements should only be interpreted by a physician or a trained health care professional who is familiar with your medical history. Through regular use of this device and recording of your measurements, you can keep your physician informed of the changes in your blood pressure. 2. Perform your measurement in a quiet place. You should be seated in a relaxed position. 3.
Product Identification LCD Display Air Connector Plug (connects to Air socket) Air Socket (insert Air connector plug here) Arm Cuff Date and Time Set Button MEMORY Recall Button INDEX START/POWER Button (on/off) Air tube Description of LCD Display Date Time Systolic Pressure Inflation/Deflation Symbol Irregular Heart Beat Symbol Diastolic Pressure Low Battery Symbol Average Measurement Symbol Memory Symbol Pulse Rate Memory Sets WHO Indicator 3 Heartbeat Symbol
Battery Installation Low battery warning: It is necessary to replace the batteries when the Low Battery symbol “ ” appears on the display, or when the display does not turn on after the POWER button is pressed. Replacing the Battery: 1. Press down on latch and lift the cover on the bottom of the monitor. 2. Insert or replace 4 x 1.5 V AA batteries into the battery compartment, ensuring to match the indicated polarity symbols. Always use new batteries. 3. Replace the battery cover.
Setting the Date and Time It is necessary to set the date and time for the unit every time batteries are initially installed or replaced. 1. While in power off mode, press the “Date and Time Set” button to enter Date and Time setting procedure and the Year value will begin to flash. 2. Press the “MEMORY Recall” button to advance the display to the desired year, press “POWER” button to confirm the year. 3. Next, the month will blink. Repeat step 2 to set the month and date, then hours, then minutes. 4.
Placement of the Pressure Sleeve It is important to avoid smoking, eating, taking medication, alcohol consumption or physical activity 30 minutes prior to taking a reading. If for any reason you are unable to or should not use your left arm, please modify the instructions for cuff application to your right arm. Your physician can help you identify which arm is best for you to take measurements from. 1. Remove any constrictive clothing or jewelry that may interfere with cuff placement. 2.
Placement of the Pressure Sleeve 8. Reaching underneath your left arm with your right hand, pull the end of the cuff towards your body to tighten the cuff. Wrap and secure the cuff, making sure that the ( ) mark remains in place as shown. 9. The cuff should fit comfortably, yet snugly around your arm. You should be able to insert one finger easily between your arm and the cuff. 10.The cuff size is suitable for use when the vertical ‘index’ mark is within the horizontal OK range.
Measurement of Pulse Rate and Blood Pressure Please read the preceding portions of this manual prior to taking your first reading. 1. Position the monitor on a flat, stable surface with the digital display in view. 2. Insert the cuff tubing connector into the port on the left side of your monitor. 3. Rest your elbow on a solid surface with your palm facing upward. Elevate your arm so that the cuff is at the same level as your heart. Relax your left hand. 4. Press the POWER button.
Measurement of Pulse Rate and Blood Pressure 9. To detect the heartbeat, the heartbeat symbol will appear and continuous flashes on the LCD display. 10.Your blood pressure measurement and pulse will display simultaneously on the screen. 11.The Hypertension Indicator will indicate your reading range on the display separately. 12. Your reading will be automatically stored in "MEMORY Recall" Button. 13. Select POWER to turn the unit off and conserve energy and battery life.
Irregular Heartbeat Detector Your digital blood pressure monitor features an Irregular Heartbeat Detector. This feature allows users to accurately monitor blood pressure even if an irregular heartbeat should occur. When an irregular heartbeat is detected, the “ ” icon will appear on the display. Note: Please consult with your physician or trained healthcare professional for further information regarding an irregular heartbeat and if this symbol appears frequently.
Memory Function Recalling Measurements in Memory: You can recall up to 60 measurements per memory bank, plus an average of last 3 currently stored measurements in memory to share with your physician or trained healthcare professional. 1. Press and release the "MEMORY Recall" Button. The unit will first display the average of last 3 currently stored measurements. 2. Continue to press the "MEMORY Recall" Button to successively view the next previously stored measurements.
Operating the Bluetooth function BPM63ZB is through the Bluetooth technology (BLE4.0) to transfer the related information of your blood pressure measurement. Please determine your device (Mobile Phone or Computer) has BLE4.0 equipped to support BPM63ZB operation. Currently, BPM63ZB can support the iOS and Android operating systems. You can also send results in memory zone as following step. 1. Press “Memory recall button” of blood pressure monitor to enter memory mode.
Troubleshooting Problem Probable Cause Recommended Action Nothing appears in the display even when the power is turned on. Low Battery Symbol appears. Batteries are drained. Replace all batteries with new ones. Battery are not correctly aligned with terminals. Reinsert batteries in the correct position. No reading after measurement. Suspicious blood pressure results. Batteries are drained. Batteries are drained. In colder temperatures batteries have weaker electrical charges.
Care and Maintenance 1. Clean the device and arm cuff carefully only with a slightly moistened cloth. 2. Do not immerse the device in water. It may cause damage if water enters. 3. Do not use any gas, strong detergent, or solvents to clean the device, including the arm cuff. 4. Disconnect the cuff and tubing from the monitor prior to storing. 5. Do not use any liquids on the monitor or cuff. 6. Use a soft, dry cloth to clean your monitor. 7. Do not store in direct sunlight, dust or excessive humidity. 8.
FCC Statement Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
EMC Tables Guidance and manufacturer's declaration-electromagnetic emissions The BPM63ZB Arm Type Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the BPM63ZB Arm Type Blood Pressure Monitor should assure that it is used in such an environment.
EMC Tables Guidance and manufacturer's declaration-electromagnetic immunity The BPM63ZB Arm Type Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the BPM63ZB Arm Type Blood Pressure Monitor should assure that it is used in such an environment.
EMC Tables Recommended separation distances between portable and mobile RF communications equipment and the BPM63ZB Arm Type Blood Pressure Monitor The BPM63ZB Arm Type Blood Pressure Monitor is intended fo use in an electromagnetic environment in which radiated RF disturbances are controlled.
Manufacturer: AViTA Corporation 9F, No.78, Sec.1, Kwang-Fu Rd., San-Chung District, 24158 New Taipei City, Taiwan No.858, Jiao Tong Road, Wujiang Economic Development Zone Jiangsu Province, P.R.C. Postcode: 215200 Made in P.R.C.