User's Manual

Operating the Bluetooth function
Measuring range :
Human Body : 34°C~43°C (93.2°F~109.4°F)
Object : 0°C~100°C (32.0°F~212.0°F)
Calibration Accuracy:
Human Body :
±0.2°C(±0.4°F): from 34 to 43 °C (93.2 to 109.4 °F)
Object : < 40°C ± 2°C; >= 40°C ± 5%
Display resolution : 0.1°C
Measuring site: Forehead
Reference body site: Ear
Operating environment :
15°C~35°C (59°F ~ 95°F)
with relative humidity up to 95% (non condensing)
Storage/ Transportation environment :
-25 to 55 °C(-13 to 131 °F)
with relative humidity up to 95% (non condensing)
Power supply : 2 x 1.5V AAA size alkaline batteries
Weight : approx. 80g (with batteries)
Dimensions :
approx. 52.4mmX44.5mmX129.9mm (L×W×H)
Operation Distance : 3 cm
For Customer Service
It is recommended that the accuracy should be checked by
manufacture every 1 years. To obtain the service please
contact AViTA Corp. for the address of the repair location.
Enclose the Proof of Purchase. Include $10.00 USD for the
return shipping and handling. Include a letter, with your
name, address, phone number, and description of the
specific problem. Pack the product carefully to prevent
damage in transit. Because of possible loss in transit, we
recommend insuring the product with return receipt
requested.
15
EMC Tables
Medical Electrical Equipment needs special precautions regarding
EMC and needs to be installed and put into service according to
the EMC information.
1. Electromagnetic/Radio Frequency Interference (EMC/RFI):
Readings may be affected if the unit is operated within a radio
frequency electromagnetic fi eld strength of approximately 3
volts per meter, but the performance of the instrument will not
be permanently affected. Care should be taken to keep the
NT16B at least 6 inches away from R/C transmitters to avoid
radio frequency interference.
2. Avoid keeping the NT16B too close to objects that continuously
generate high heat (like a hot plate) for long periods of time,
which can cause overheating of the NT16B.
Table 1 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic emissions
The NT16B TOUCH FREE Infrared Thermometer is intended for use in the
electromagnetic environment specified below. The customer or the user of
the NT16B TOUCH FREE Infrared Thermometer should assure that it is
used in such an environment.
Emissions test Compliance
Electromagnetic
environment-guidance
RF emissions CISPR 11 Group 1 The NT16B TOUCH FREE
Infrared Thermometer uses RF
energy only for its internal
function.
Therefore, its RF emissions are
very low and are not
likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
Flicker emissions
IEC 61000-3-3
Not applicable
The NT16B TOUCH FREE
Infrared Thermometer is
suitable for use in all
establishments other that
domestic and those directly
connected to the public
low-voltage power supply
network that supplies buildings
used for domestic purposes.
16