Please read this manual thoroughly and ensure all contents are fully understood before using the apparatus. Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the tester or to the equipment under test, adhere to the following rules: Before using the tester inspect the case. Do not use the tester if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic.
Replace the battery as soon as the battery indicator appears. With a low battery, the meter may produce false readings that can lead to electric shock and personal injury. Remove the connection between the testing leads and the circuit being tested, and turn the meter power off before opening the meter case. The internal circuit of the meter shall not be altered at will to avoid damage of the meter and any accident.
MTB01 & MTB02 User Manual
1) 2) 3) 4) 5) 6) LCD display Range switch hFE socket 10A jack VΩmA jack COM jack Multitester comparison table Model MTB01 MTB02 DCV ACV DCA Ω Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes N\A hFE Yes Yes Yes N\A Technical Specifications Accuracies are guaranteed for 1 year, 23ºC ± 5ºC, less than 80% RH. DC Voltage RANGE 200mV 2000mV 20V 200V 500V RESOLUTION 100uV 1mV 10mV 100mV 1V ACCURACY ± (0.5% of rdg + 3D) ± (0.8% of rdg + 5D) ± (1.
AC Voltage RANGE 200V 500V RESOLUTION 100mV 1V ACCURACY ± (2.0% of rdg + 10D) ± (2.0% of rdg + 10D) RESPONSE: Average response calibrated in rms of a sine wave FREQUENCY RANGE: 45Hz - 450Hz OVERLOAD PROTECTION: 500Vdc or 500Vrms for all ranges Audible Continuity RANGE DESCRIPTION Built-in buzzer sounds if resistance is less than 30±20Ω OVERLOAD PROTECTION: 15 seconds maximum 220Vrms DC Current RANGE 200uA 2000uA 20mA 200mA 5A RESOLUTION 100nA 1uA 10uA 100uA 10mA ACCURACY ± (1.8% of rdg + 2D) ± (2.
Resistance RANGE 200Ω 2000Ω 20kΩ 200kΩ 2000kΩ RESOLUTION 0.1Ω 1Ω 10Ω 100Ω 1kΩ ACCURACY ± (1.0% of rdg + 10D) ± (1.0% of rdg + 4D) MAXIMUM OPEN CIRCUIT VOLTAGE: 3V OVERLOAD PROTECTION: 15 seconds maximum 220Vrms OPERATING INSTRUCTIONS DC & AC VOLTAGE MEASUREMENT 1. Connect red test lead to “VΩmA” jack, black lead to “COM” jack. 2.
2. Set RANGE switch to desired DCA position. 3. Open the circuit to be measured, and connect test leads in SERIES with the load in with current is to measure. 4. Read current value on Digital Display. 5. Additionally, “5A” function is designed for intermittent use only. Maximum contact time of the test leads with the circuit is 10 seconds, with a minimum intermission time of 15 minutes between tests. RESISTANCE MEASUREMENT 1. Red lead to “VΩmA”. Black lead to “COM”. 2.
TRANSISTOR hFE MEASUREMENT 1. Set RANGE switch to the hFE position. 2. Determine whether the transistor is PNP of NPN type and locate the Emitter, Base and Collector leads. Insert the leads into the correct holes of the hFE Socket. 3. The meter will display the approximate hFE value at the condition of base current 10μA and VCE2.8V. AUDIBLE CONTINUITY TEST 1. Red lead to “VΩmA”, Black lead to “COM”. 2. RANGE switch to “ ” position. 3. Connect test leads to two points of circuit to be tested.
ACCESSORIES Instruction manual Set of test leads (red and black) 9V PP3 battery EN61010–1:2010 This product is classed as Electrical or Electronic equipment and should not be disposed with other household or commercial waste at the end of its useful life. The goods must be disposed of according to your local council guidelines. Errors and omissions excepted. Copyright© 2014. AVSL Group Ltd.