Triggering
6
6.2 Setting the Trigger Delay
<For a description of this function, see page 1-13>
Relevant Keys
ACTION DELAY
X - Y
MENU MENU
PHASE
SEARCH
HORIZONTAL
VERTICAL
TRIGGER
TRIG D
HELP
ESC
SELECTRESET
SIMPLE
MODE POSITION
MATH
MISCFILE
CLEAR
TRACE
HISTORY
MEASURE
SETUP
PRESET
DISPLAY
CURSOR
CH
1
CH
2
CH
3
CH
4
POWER
COPY
IMAGE SAVE
SHIFT
ZOOM
GO/NO-GO
ENHANCED
ACQSTART/STOP
SNAP
SHOT
VDIV TIME DIV
Operating Procedure
1. Press SHIFT to set the keys in the shifted condition.
Functions marked in purple on the panel become active.
2. Press POSITION.
3. Units, ms, us, ns, and ps are displayed on the screen. Press the soft key
corresponding to the desired unit.
Turn the jog shuttle to set the trigger delay. You can move between the digits
using the arrow keys. Pressing RESET resets delay to 0 s.
You can set the trigger delay by pressing the Delay soft key and turning the jog
shuttle.
Explanation
Although the display usually shows the waveform before and after the trigger point, using
the delay function, it is possible to display the acquired waveform after a fixed time
period elapses.
Setting Range for Trigger Delay
0 to 4 s (Resolution is 1 ÷ sample rate)
Delay Time
Trigger Point
T(Trigger Position)
Points for Attention
When T/div is changed, the trigger delay remains unchanged. However, its position in
relation to the overall record length changes.
If the timebase is provided by an external clock, the trigger delay is fixed at 0s (no
delay).