7.2 Acquisition Mode
<For a description of this function, see page 1-15>
Relevant Keys
ACTION DELAY
X - Y
MENU MENU
PHASE
SEARCH
HORIZONTAL
VERTICAL
TRIGGER
TRIG D
HELP
ESC
SELECTRESET
SIMPLE
MODE POSITION
MATH
MISCFILE
CLEAR
TRACE
HISTORY
MEASURE
SETUP
PRESET
DISPLAY
CURSOR
CH
1
CH
2
CH
3
CH
4
POWER
COPY
IMAGE SAVE
SHIFT
ZOOM
GO/NO-GO
ENHANCED
ACQSTART/STOP
SNAP
SHOT
VDIV TIME DIV
Operating Procedure
Setting the Acquisition Mode
1. Press ACQ.
2. Press the Mode soft key to display the acquisition mode menu.
3. Press the soft key corresponding to the desired mode from Normal, Envelope, or
Average.
You may not be able to select some modes depending on the record length and
sample rate settings.
Setting the Acquisition Count
4. Press the Count soft key.
This is not available when the trigger mode is Single or Single (N).
5. Turn the jog shuttle to set the acquisition count.
You can use the arrow keys to move between the digits.
In addition, pressing RESET sets the count to Infinite.
When the acquisition mode is Average and you have selected Infinite, go to step 6.
Setting the Attenuation (When the Acquisition Mode is Average and the Count is
Set to Infinite)
6. Press the Weight soft key.
7. Turn the jog shuttle to set the attenuation.