User`s manual

9-10
IM 701610-01E
For H&V Cursors (When Displaying the X-Y Waveform)
Selecting the waveform to be measured
See steps 1-3 to set the Type to H&V.
4. Press the Trace soft key and select the target waveform.
Moving the Cursor
5. Press the Curs1
/Curs2 soft key to activate the jog shuttle on Curs1 (Cursor 1).
6. Turn the jog shuttle to move Curs1 (Cursor 1).
7. Move Curs2 (Cursor 2) in the same manner. If you activate the jog shuttle on both
Curs1 and Curs2, you can move both cursors at the same time.
8. Move Curs1 /Curs2 in the same manner according to steps 5 through 7.
Cursor1
Cursor2
Cursor1
Cursor2
Explanation
Restrictions
Cursor measurements cannot be made on the following waveforms.
Snapshot waveforms
Accumulated waveforms other than the newest waveform
Cursor Types and Measurement Items: Type (When Not Displaying the X-Y Waveform)
•H (Horizontal) Cursor
Measures the Y-axis value at the cursor.
Y1: The Y-axis value at Cursor1
Y2: The Y-axis value at Cursor2
Y: The difference between the Y-axis values at Cursor1 and Cursor2
•V (Vertical) Cursor
Measures the X-axis value at the cursor. Also choose the order of the data.
X1: The X-axis value at Cursor1
X2: The X-axis value at Cursor2
X : The difference between the X-axis values at Cursor1 and Cursor2
1/X: The inverse of the difference between the X-axis values at Cursor1 and Cursor2
Y1: The Y-axis value at Cursor1
Y2: The Y-axis value at Cursor2
Y: The difference between the Y-axis values at Cursor1 and Cursor2
9.1 Measuring Waveforms Using Cursors