User`s manual

9-34
IM 701610-01E
9.4 Performing Automated Measurements of Waveform Parameters on Dual Areas
Precautions to be Taken when Performing Automated Measurement of Waveform
Parameters
The measurement value displays “*****” if the measurement is not possible.
For waveforms of small amplitude, correct measurements may not be possible.
When there are two or more periods of waveform in the measurement range, the time
axis parameter measurement is performed on the first cycle.
Every time the waveform is updated, the waveform parameters are measured
automatically.
When loading waveform data of the maximum record length, automatic measurement
of waveform parameters cannot be performed on calculated waveforms. ***** is
displayed for the measured results.