User`s manual
1-9
IM 701610-01E
Functions
1
A Delay B Trigger (Enhanced Trigger) → Section 6.9
This function activates a trigger the first time condition B becomes true after condition A
becomes true and the specified time elapses.
L
HL L
H
HHHHLLL
H
CH1
CH2
CH1
CH2
Trigger
LH HL
L
Pattern A is True Pattern B is True
1 µs
When pattern A: CH1 = L, CH2 = L, Enter, When pattern B: CH1 = H, CH2 = H, Enter, Delay = 1 µs
Pattern Trigger (Enhanced Trigger) → Section 6.10
Multiple trigger sources are selected, and a trigger is activated when all of the trigger
conditions set for each trigger source become true or false. Trigger conditions are
established by setting combinations of the state (High or Low) of each trigger source.
Furthermore, one of the trigger sources can be used as the clock signal, and triggering is
synchronized with this clock signal.
LH L L L
H
HHHHLLLL
L
HLLH
HLL
H
CH1
CH2
CH3*
CH4*
CH1
CH2
CH3*
CH4*
Trigger is activated.
Example: a trigger is activated
when CH1: L, CH2: L, CH3*: H and CH4*: L(* The DL1620 is not equipped
with channels 3 and 4.)
Pulse Width Trigger → Section 6.11
The time period during which the specified condition is met or not met is compared with
the specified time period. The trigger condition is set with the AND of the signal state of
each channel (High, Low, or Don’t Care) or the AND of the window conditions of each
channel (IN, OUT, or Don’t Care).
H
L
L
H
H
H
H
LL
L
H
CH1
CH2
CH1
CH2
LH
H
LL
500 ns
300 ns
400 ns
A
B
C
D
450 ns
1.3 Setting a Trigger