Waveform Analysis
9
Connecting to Another Instrument
CAUTION
• Never apply an external voltage to the NO-GO OUT and GO OUT
terminals, otherwise damage to the instrument may result.
• When connecting the GO/NO-GO signal output externally, make sure
not to connect other signal pins by mistake. Errors in connection may
cause damage to this instrument or to the other connected instrument.
• Do not mistakenly insert the USB cable to the GO/NO-GO output
terminal. This may cause damage to the DL1620/DL1640/DL1640L.
To connect the DL1620/DL1640/DL1640L to external equipment, use the GO/NO-GO
cable (366973, sold separately).
Do not use the GO/NO-GO cable (366973, sold separately) for any other purpose than
for carrying out GO/NO-GO determination on the DL1620/DL1640/DL1640L.
• About the GO/NO-GO Cable (366973, Sold Separately)
1
6
To the DL1640/DL1640L
Approx. 1.5 m
RJ-12 Connector
Wires are separated.
Modify them to suit your needs.
Pin No. Signal Name Logic
3
4
5
GO OUT
NO-GO OUT
Negative
Negative
GND
Color
White
Green
Blue
2NC
Yellow
9.11 Using the GO/NO-GO Signal Output Function