User`s manual
1-18
IM 701610-01E
Waveform Labels → Section 8.5
You can assign an arbitrary label (up to 8 characters) to each waveform.
Waveform Label
Trigger Mark
Scale Values
Selects use of
2 waveform windows.
Display Interpolation <Section 8.2>
This feature selects the type of interpolation applied in areas where there are less than
500 sample points (Less than 250 points in the zoom window when zooming on the
waveform using Main & Z1 & Z2) per 10 time axis divisions. (These areas are referred to
as interpolation areas.) Three settings are available.
Line Interpolation
Interpolates between two dots using a straight line.
Sine Interpolation
Generates interpolation data using the function
sin x
x
then interpolates between two dots
using resulting sine curve. Sine interpolation is suitable for observation of sine waves.
Pulse Interpolation
Interpolates between two dots using a step.
No Interpolation
Displays measurements as discrete dots, without interpolation.
1.4 Setting the Acquisition and Display Conditions