User`s manual
15-6
IM 701610-01E
Code Message Corrective Action Reference Section
920 Cannot set the acquisition mode to Average Change the acquisition mode or the trigger 6.1, 7.2
when the trigger mode is set to Single or mode.
Single(N).
921 It is not possible to make a setting that will Setting or set the trigger mode to a different 6.1
result in the repetitive mode when the trigger mode.
mode is set to Single (N).
923 Cannot set the trigger mode to Single or Change the acquisition mode or the trigger 6.1, 7.2
Single(N) when the acquisition mode is mode.
Average.
926 Cannot set the trigger mode to Single(N) Turn OFF the repetitive sampling mode, 5.12, 7.1
during repetitive sampling mode. lower the T/div setting, or shorten the
record length.
927 Cannot set this parameter during repetitive Turn OFF the repetitive sampling mode, lower 5.12, 7.1
sampling mode. the T/div setting, or shorten the record length.
930 Not possible in the current record length. Change the record length. 7.1
931 Cannot carry out computation at the current Change the record length. 7.1
record length.
932 The operation is not possible when Unload the files. 11.6
waveforms are loaded.
Unload the loaded files from the FILE menu.
933 Setting or executing is not possible during Abort the search. 7.7, 7.8
the search operation.
934 Search pattern does not exist. Change the search conditions. 8.9
Execute the search.
935 Settings cannot be changed or executed Abort the search. 7.7, 7.8
during the history search operation.
936 The record cannot be selected Check the record number using Show Map. 7.7, 7.8
937 History record does not exist. History record is not created in the average 7.2
mode, repetitive sampling mode, or roll mode.
938 Setting or executing is not possible during Set the history Display Mode to One. 7.6
FFT recalculation.
Abort the operation by setting the history
Display Mode to One.
940 Cannot be configured or executed while Set the history Display Mode to One. 7.6
updating the history all display.
Aborted when history display mode is set to
One.
941 Cannot output color in this format. Turn OFF the color. 10.3
942 Zones cannot be edited in the following cases: Display the main window and the target 9.9
• When the main window is not displayed. waveform.
• When the target waveform is not displayed.
943 The zone waveform does not exist. Create the zone waveform. 9.9
944 The zone is being edited. Select Quit to exit zone editing. 9.9
To perform other operations, select Quit to
exit zone editing.
945 Zones determination is not possible in the Display the main window and the target window 9.9
following cases: and create the zone waveform.
• When the main window is not displayed.
• When the target waveform is not displayed.
• When the zone waveform does not exist.
948 Processing statistics. Abort statistical processing. 9.3
To perform other operations, abort the
statistical processing.
949 Settings cannot be entered in the current — 6.1
trigger mode.
Set the trigger mode to Single.
950 Setting not allowed. Invalid byte or bit. — —
951 Cannot be set when CS channels are not — —
specified.