System Operation Errors (960 to 968, 972 to 981)
Code Message Corrective Action Reference Section
952 Cannot set this parameter when the 7.2
acquisition mode is Average.
961 Failed to backup setup data. Check the condition of the built-in lithium
Will initialize. battery using the overview screen.
Servicing is required.
966 Fan stopped ; Turn OFF the power immediately.
Turn off the power immediately Servicing is required.
967 Backup battery is flat Servicing is required to replace the battery.
972 Fatal error in the communication driver Servicing is required.
982 The USB devices power consumption 16.8
exceeded the capacity of the USB hub.
983 The voltage of power supply battery is too low. Check the output from the DC power 3.3
Turn off the power immediately. supply being used.
Please do charge of the battery or replace.
984 The voltage of power supply battery is too high. Check the output from the DC power 3.3
Turn off the power immediately. supply being used.
Confirm the voltage of the battery.
Note
When servicing is required, double-check by initializing the instrument.