Functions
1
Statistical Processing for Each Period
Divides the displayed waveform using a period that is automatically calculated and
performs statistical processing on the measured values over the determined period.
Statistical processing is performed from the oldest measured data of the displayed
waveform.
CH2
CH1
a
b
c
The parameters for automated measurement are measured in
each range a, b, and c, and statistical processing is performed
on each automated measurement parameter in the order a, b,
and c.
Automated measurement parameters of other channels are also
measured over each range a, b, and c.
Automated measurement can also be performed on the period
of each waveform.
Example in which CH1 is selected as the target waveform for determining the period
Statistical Processing of Historical Data
Performs automated measurement of waveform parameters on the waveform that is
acquired using the history memory function and performs statistical processing.
Statistical processing is performed from the oldest waveform.
Perform Automated Measurements of Waveform Parameters on Dual Areas
Section 9.4
You can specify two areas and perform automated measurement of waveform
parameters on each area. You can also perform computation on the parameters
determined in the two areas. Statistical processing for each period is not possible.
1.5 Analyzing the Waveform