User`s manual

3-12
IM 701610-01E
3.5 Compensating the Probe (Phase Correction)
CAUTION
Never apply an external voltage to the COMP terminal, as damage to
the instrument may result.
Operating Procedure
1. Turn ON the power switch.
2. Connect the probe to the input terminal to which the signal is to be applied.
3. Touch the probe’s tip against the probe compensation signal output terminal and
connect the grounding wire to the functional earth terminal.
4. Perform auto setup using the procedure described in section 4.3.
5. Insert a screwdriver into the phase adjusting hole in the probe and turn the trimmer
so that the displayed waveform becomes square.
Functional Earth Terminal
Phase Adjusting Hole
Probe Compensation (COMP)
Signal Output Terminal