User`s manual

5-3
IM 701610-01E
Vertical and Horizontal Axes
5
5.2 Setting the Vertical Position of a Waveform
Explanation
Range of Movement
The vertical position can be moved in the range between ±4 div from the center position
in the waveform display frame.
Setting Resolution
0.01 div
Confirming the Vertical Position
For input waveforms and computed waveforms, the ground level and vertical position are
marked on the left of the waveform display frame.
500 mV/div, Offset: –1 V, Offset Cancel: OFF, Position: 0 div
Ground Level
Mark
Vertical Position
Mark
Note
The data which go out of the waveform display frame from moving the vertical position are
handled as overflow data.
If the display waveform goes out of the waveform display frame from moving the vertical
position during the waveform acquisition is starting, a chopped waveform is displayed as
shown in the following figure even if the vertical position is returned to its original position
after stopping the acquisition.
If the vertical position is moved, the effective data range also changes. For details, see
section 1.2.