Technical data

2400 Series Microwave Synthesizer: Remote Program Examples
C-2 2400 Series Operation Manual, 33237 Revision A, May 2004
C.1.3 Simple List Mode
C.1.4 Comprehensive List Mode
Sequence Command Description
1 LIST:SEQ:AUTO ON Activate list mode.
2LIST:FREQ
5000000000.0,5000000000.0,5000000000.0
Add 3 list points to a list with frequency 5 GHz.
3 LIST:POW 8.000,5.000,0.000 Set power to 8, 5 and 0 dBm separately.
4 LIST:DWEL 0.200000, 0.200000, 0.200000 Set dwell(step) time to 0.2 second.
5 LIST:PRECompute? 1 Pre-compute the created list data
6 LIST:REPeat SWEEP Set list repeat type to single sweep
7 TRIGger:SOURce BUS Set list trigger mode to GPIB (software trigger)
8OUTP ON Turn RF on.
9 *TRG Trigger the list.
Sequence Command Description
1 LIST:SEQ:AUTO ON Activate list mode
2 LIST:FREQ 1000000000.0,2000000000.0,3000000000.0 Add 3 list points to a list with frequency 1, 2
and 3 GHz separately
3 LIST:POW 0.000,0.000,0.000 Set power to 0 dBm for 3 points.
4 LIST:DWEL 0.250000, 0.250000, 0.250000 Set dwell (step) times to 0.25 second
5 LIST:PRECompute? 1 Pre-compute the created list data
6 LIST:REPeat CONT Set list repeat type to continuous
7 TRIGger:SOURce BUS Set list trigger mode to GPIB (software
trigger)
8OUTP ON Turn RF on
9 *TRG Trigger the list
10 LIST:DEL:LIST 1 Clear the existing list
11 LIST:FREQ 6000000000.0,6000000000.0,6000000000 Add 3 list points to a list with frequency 6
GHz.