Technical data

2400 Series Microwave Synthesizer: Remote Program Examples
C-4 2400 Series Operation Manual, 33237 Revision A, May 2004
C.2.3 Simple List Mode
Sequence Command Description
9 LR 1 Activate the list mode.
10 LA 1 0 Add the first list point to a list.
11 LF 1 1 10.000000 GHZ Set the frequency of the first list point to 10 GHz.
12 LL 1 1 0.000000 DBM Set the power output of the first list point to 0
dBm.
13 LT 1 1 0.250000 S Set the dwell time of the first list point to 0.25
second.
14 LA 1 1 Add the second list point to the list