Technical data

Chapter 1:
1-8 2400 Series Operation Manual, 33237 Revision A, May 2004
* L - 2400L series, M - 2400M series, AL - 2400AL series, AM - 2400AM series
RAMP OUT L, M,(Option 43) 0 to 10 volt ramp output scaled to frequency sweep.
AM IN
M, AM
External AM input 600
. Any waveform compatible with
bandwidth considerations. 1 Vp-p for 50% depth.
FM IN
M, AM
External FM 50
input. Any waveform compatible with
bandwidth considerations. 1 Vp input for maximum
deviation.
PM IN
M, AM
+5.0 volt, 50 input.
PM TRIGGER IN
M, AM
+5.0 volt, 50
input. Triggers a delayed pulse output
specified in Internal PM menu (Option 24 required).
AM OUT
M, AM (Option 24)
Modulation generator output, 2 Vp-p into 10 k
.
FM OUT
M, AM (Option 24)
Modulation generator output, 2 Vp-p into 10 k .
PM OUT
M, AM (Option 24)
+4.0 volt video representation of the pulsed signal output
at 50 .
PM SYNC OUT
M, AM (Option 24)
50 ns width, TTL level output into 50
. Coincident with
Pulse trigger event + 30 ns.
GPIB Interface
L, M, AL, AM
IEEE STD 488.2 General purpose instrument bus.
RS-232
L, M, AL, AM
Serial Interface, DB9 connector.
AC Power Input
L, M, AL, AM
90-253 VAC, auto-sensing, 47 Hz to 64 Hz.
Table 1-2: 2400 Series Rear Panel I/O Connector Descriptions
I/O Connector
Name
2400 Series Availability Description