Technical data
Chapter 2: 2400 Operation
2-8 2400 Series Operation Manual, 33237 Revision A, May 2004
2.3.6.1.5 Resolution
Press the Resolution softkey. Using the step keys or rotary knob to enter the ramp fre-
quency sweep resolution. The settable range is 401, 801 and 1601.
2.3.6.1.6 Ramp Power Menu
Press the Ramp Power Menu softkey to enter the Ramp Power Sweep menu.
2.3.7 Configure a Ramp Power Sweep
For ramp power sweep operation, Option 43 must be installed. Press the RAMP button to access the
RAMP menu, and then press the Ramp Power Menu softkey to enter the Ramp Power menu. If option
43 is not installed, a status message will appear at the bottom of the display indicating that option 43
is not installed. After a ramp calculation is performed by the 2400, the unit will continuously sweep.
There is no provision for an externally triggered ramp sweep. The maximum settable range for ramp
sweep is 45 dB. If option 26 is installed in the 2400, the ramp power sweep operation prevents the
attenuator from switching during a ramp power sweep. For power sweeps with levels below -20 dBm,
it is necessary to set the step attenuator before setting the start and stop power sweep settings.
Illustration 2.5 | Ramp Power Menu
2.3.7.1 Ramp Power Sweep Parameters
2.3.7.1.1 Start Power
Press the Start Power softkey. Use the numeric keypad, step keys or rotary knob to
enter the starting power level.