Technical data

Chapter 2: 2400 Operation
2-10 2400 Series Operation Manual, 33237 Revision A, May 2004
units key to recall the set up. After a successful recall, the display menu will remain in the System 1
Menu.
Contrast Press the Contrast softkey to modify the display contrast. The display contrast can be modi-
fied using the Step Up/Step Down keys, numeric keypad and Rotary Knob. The contrast setting range
is 1 to 15
Sound To enable or disable the system sound, press the Sound softkey to select the Sound parame-
ter. The sound setting is toggled using the Step Up/Step Down Keys. The available selections are On
and Mute. Available system sounds are the increment/decrement functions for the Rotary Knob and
System Operational Error Notification.
Attenuation This selection provides control of the system step attenuator if it is installed in the 2400
(Option 26). Press the Attenuation softkey to modify the attenuator setting. Using the Step Up/Step
Down keys, set the attenuator to automatically adjust or manual control in 10 dB steps to 90 dB.
Illustration 2.6 | System Menu 1
2.3.8.2 System Menu 2
GPIB Address Press the GPIB Address to move the cursor to the GIPB address parameter. The GPIB
address can be set from 1 to 30 using the numeric keypad, Step Up/Step Down keys or Rotary Knob.
Language Interface Press the Language Interface softkey to move the cursor to the Interface param-
eter. Select SCPI or GT12000 selection to set the remote command language using the Step Up/Step
Down keys. SCPI is Standard Commands for Programmable Instruments. GT12000 is Giga-tronics
native command set. See Giga-tronics 12000A Operations Manual for the remote command language
description.
General Information This field displays the 2400 system information identifying the model type,
firmware version, serial number and firmware build date information.
SYSTEM 1
OFS SLP
To Register (0-9)
No register has been saved
8
ON
Save
Recall
Sound
Contrast
Option 26 not installed
Attenuation
System 2
Menu