Technical data
Chapter 2: 2400 Operation
2-12 2400 Series Operation Manual, 33237 Revision A, May 2004
Illustration 2.8 | Amplitude Modulation Menu 1
2.3.9.1.1 Internal Waveform Source (AM Menu 2)
The internal AM requires setting the following parameters: Depth, Rate and Waveform.
• Press the Depth softkey to move the cursor to the Depth parameter. Edit the depth parameter using
the numeric keypad, step up/step down keys or rotary knob. The maximum depth setting is 95%.
• Press the Rate softkey to move the cursor to the AM Rate parameter. Edit the AM rate parameter
using the numeric keypad, step up/step down keys or rotary knob.
• Press the Waveform softkey to move the cursor to the AM Waveform parameter. Change the wave-
form by pressing the Step Up or Step Down keys. The available selections are Sine, Triangle, Ramp or
Square.
Illustration 2.9 | Amplitude Modulation Menu 2
2.3.9.1.1 Internal Noise Source (AM Menu 3)
Press the Depth softkey to move the cursor to the Depth parameter. Edit the depth parameter using
the numeric keypad, step up/step down keys or rotary knob. The maximum depth setting is 95%.
AM MENU 1: External Source
On/Off
0.00% / V
AM
Sensitivity
AM 2/3
AM MENU 2: Internal Waveform
On/Off
Sine/Triangle /Ramp/Square/Noise
AM
Rate
AM 3/3
Depth
90.00 %
1.00 kHz
Waveform