Technical data

Chapter 4: Specification & Performance Verification
4-2 2400 Series Operation Manual, 33237 Revision A, May 2004
4.1.1.2 Resolution
0.1 Hz
4.1.1.3 Accuracy & Stability (Identical to Timebase Oscillator)
4.1.2 RF Output Power
4.1.2.1 Maximum Leveled Output
2,3
(0 to 35°C)
2
Specification applies over the 0 to 35°C range and degrades <2.0 dB from 35 to 55°C.
3
Step attenuator (option 26) reduces power by 1.5 dB to 20 GHz and 2.0 dB above 20 GHz
4
20 - 26.5 GHz for model 2426
10 8.0 - 15.99 GHz 1/2
11 16.0 - 31.99 GHz 1/4
12 32.0 - 40.00 GHz 1/8
4.1.1.3.1 Timebase (Internal) 10 MHz
4.1.1.3.2 Aging Rate
(after 30 minutes warm up time)
< 5 x 10
-10
/day, 2400L and 2400M Series only
<1 x 10-8/day. 2400A Series only
4.1.1.3.3 Temperature Stability ± 5 x 10
-10
/°C, 2400L and 2400M Series only
±2 x 10
-8
/°C, 2400A Series only
4.1.1.3.4 10 MHz Reference Output TTL into 50
4.1.1.3.5 External Reference Input 10 MHz or 100 MHz ± 1 ppm
> - 5 dBm into 50
Frequency Range (GHz) .01 - 8 GHz 8 - 20 GHz
20 -40 GHz
4
20 GHz Model + 16 +15
26.5 GHz Model +13 +9 +10
40 GHz Model +13 +9 +9
Band Frequency N